TY - JOUR
T1 - Safety profiles of percutaneous left atrial appendage closure devices
T2 - An analysis of the Food and Drug Administration Manufacturer and User Facility Device Experience (MAUDE) database from 2009 to 2016
AU - Jazayeri, Mohammad Ali
AU - Vuddanda, Venkat
AU - Turagam, Mohit K.
AU - Parikh, Valay
AU - Lavu, Madhav
AU - Atkins, Donita
AU - Earnest, Matthew
AU - Di Biase, Luigi
AU - Natale, Andrea
AU - Wilber, David
AU - Reddy, Yeruva Madhu
AU - Lakkireddy, Dhanunjaya R.
N1 - Publisher Copyright:
© 2017 Wiley Periodicals, Inc.
PY - 2018/1
Y1 - 2018/1
N2 - Background: Percutaneous left atrial appendage closure (LAAC) is a viable option for AF patients who are unable to tolerate long-term oral anticoagulation (OAC). Objective: We sought to assess the safety of two commonly used percutaneous devices for LAA closure in the United States by analysis of surveillance data from the FDA Manufacturer and User Facility Device Experience (MAUDE) database. Methods: The MAUDE database was queried between May 1, 2006 and May 1, 2016 for LARIAT® (SentreHEART Inc., Redwood City, CA, USA) and WATCHMAN™ (Boston Scientific Corp., Marlborough, MA, USA) devices. Among 622 retrieved medical device reports, 356 unique and relevant reports were analyzed. The cumulative incidence of safety events was calculated over the study period and compared between the two devices. Results: LAAC was performed with LARIAT in 4,889 cases. WATCHMAN was implanted in 2,027 patients prior to FDA approval in March 2015 and 3,822 patients postapproval. The composite outcome of stroke/TIA, pericardiocentesis, cardiac surgery, and death occurred more frequently with WATCHMAN (cumulative incidence, 1.93% vs. 1.15%; P = 0.001). The same phenomenon was observed when comparing the WATCHMAN pre- and postapproval experiences for the composite outcome, as well as device embolization, cardiac surgery, and myocardial infarction. Conclusions: MAUDE-reported data show that postapproval, new technology adoption is fraught with increased complications. Improved collaboration between operators, device manufacturers, and regulators can better serve patients through increased transparency and practical postmarket training and monitoring mechanisms.
AB - Background: Percutaneous left atrial appendage closure (LAAC) is a viable option for AF patients who are unable to tolerate long-term oral anticoagulation (OAC). Objective: We sought to assess the safety of two commonly used percutaneous devices for LAA closure in the United States by analysis of surveillance data from the FDA Manufacturer and User Facility Device Experience (MAUDE) database. Methods: The MAUDE database was queried between May 1, 2006 and May 1, 2016 for LARIAT® (SentreHEART Inc., Redwood City, CA, USA) and WATCHMAN™ (Boston Scientific Corp., Marlborough, MA, USA) devices. Among 622 retrieved medical device reports, 356 unique and relevant reports were analyzed. The cumulative incidence of safety events was calculated over the study period and compared between the two devices. Results: LAAC was performed with LARIAT in 4,889 cases. WATCHMAN was implanted in 2,027 patients prior to FDA approval in March 2015 and 3,822 patients postapproval. The composite outcome of stroke/TIA, pericardiocentesis, cardiac surgery, and death occurred more frequently with WATCHMAN (cumulative incidence, 1.93% vs. 1.15%; P = 0.001). The same phenomenon was observed when comparing the WATCHMAN pre- and postapproval experiences for the composite outcome, as well as device embolization, cardiac surgery, and myocardial infarction. Conclusions: MAUDE-reported data show that postapproval, new technology adoption is fraught with increased complications. Improved collaboration between operators, device manufacturers, and regulators can better serve patients through increased transparency and practical postmarket training and monitoring mechanisms.
KW - Food and Drug Administration
KW - LARIAT
KW - WATCHMAN
KW - air embolism
KW - atrial fibrillation
KW - cardiac surgery
KW - left atrial appendage closure
KW - left atrial appendage occlusion
KW - malfunction
KW - mortality
KW - outcomes
KW - pericardial effusion
KW - stroke
KW - transient ischemic attack
UR - http://www.scopus.com/inward/record.url?scp=85033221133&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85033221133&partnerID=8YFLogxK
U2 - 10.1111/jce.13362
DO - 10.1111/jce.13362
M3 - Article
C2 - 28988455
AN - SCOPUS:85033221133
SN - 1045-3873
VL - 29
SP - 5
EP - 13
JO - Journal of Cardiovascular Electrophysiology
JF - Journal of Cardiovascular Electrophysiology
IS - 1