Abstract
To explore the possibility of using Compton scattering method to characterize the energy spectra of clinical beams. This method is experimentally tested on a Co60 radiation therapy machine and independently verified using Monte Carlo (MCNP4B) simulation. Measured spectrum and MCNP simulated spectrum including detector response for 90 degree scattered beam show good agreement.
Original language | English (US) |
---|---|
Pages (from-to) | 1082-1084 |
Number of pages | 3 |
Journal | Annual International Conference of the IEEE Engineering in Medicine and Biology - Proceedings |
Volume | 2 |
DOIs | |
State | Published - 2000 |
Externally published | Yes |
ASJC Scopus subject areas
- Signal Processing
- Biomedical Engineering
- Computer Vision and Pattern Recognition
- Health Informatics