Possible errors in uroflowmetry

Matthew Lief, Zafar Khan, Elliot Leiter, Arnold Melman

Research output: Contribution to journalArticlepeer-review

7 Scopus citations

Abstract

The possibility of alteration of uroflow rates by rapid bladder filling was evaluted in 61 consecutive patients referred for urodynamic evaluation. Nine of the 28 patients who were able to void did in fact demonstrate a significant reduction in uroflow rates after the bladder had been rapidly filled. Thus, the manner of obtaining the uroflow should be taken into account when making a clinical decision based on that test.

Original languageEnglish (US)
Pages (from-to)179-183
Number of pages5
JournalNeurourology and Urodynamics
Volume3
Issue number3
DOIs
StatePublished - 1984
Externally publishedYes

Keywords

  • bladder fatigue
  • cystometrogram
  • fill‐pull
  • plastoelastic
  • uroflowmetry