In vivo real-time rectal wall dosimetry for prostate radiotherapy

Nicholas Hardcastle, Dean L. Cutajar, Peter E. Metcalfe, Michael L F Lerch, Vladimir L. Perevertaylo, Wolfgang A. Tome, Anatoly B. Rosenfeld

Research output: Contribution to journalArticle

38 Citations (Scopus)

Abstract

Rectal balloons are used in external beam prostate radiotherapy to provide reproducible anatomy and rectal dose reductions. This is an investigation into the combination of a MOSFET radiation detector with a rectal balloon for real-time in vivo rectal wall dosimetry. The MOSFET used in the study is a radiation detector that provides a water equivalent depth of measurement of 70 m. Two MOSFETs were combined in a face-to-face orientation. The reproducibility, sensitivity and angular dependence were measured for the dual MOSFET in a 6 MV photon beam. The dual MOSFET was combined with a rectal balloon and irradiated with hypothetical prostate treatments in a phantom. The anterior rectal wall dose was measured in real time and compared with the planning system calculated dose. The dual MOSFET showed angular dependence within ±2.5% in the azimuth and +2.5%/-4% in the polar axes. When compared with an ion chamber measurement in a phantom, the dual MOSFET agreed within 2.5% for a range of radiation path lengths and incident angles. The dual MOSFET had reproducible sensitivity for fraction sizes of 2-10 Gy. For the hypothetical prostate treatments the measured anterior rectal wall dose was 2.6 and 3.2% lower than the calculated dose for 3DCRT and IMRT plans. This was expected due to limitations of the dose calculation method used at the balloon cavity interface. A dual MOSFET combined with a commercial rectal balloon was shown to provide reproducible measurements of the anterior rectal wall dose in real time. The measured anterior rectal wall dose agreed with the expected dose from the treatment plan for 3DCRT and IMRT plans. The dual MOSFET could be read out in real time during the irradiation, providing the capability for real-time dose monitoring of the rectal wall dose during treatment.

Original languageEnglish (US)
Pages (from-to)3859-3871
Number of pages13
JournalPhysics in Medicine and Biology
Volume55
Issue number13
DOIs
StatePublished - Jul 7 2010
Externally publishedYes

Fingerprint

Prostate
Radiotherapy
Radiation
Photons
Anatomy
Ions
Water
Therapeutics

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

Hardcastle, N., Cutajar, D. L., Metcalfe, P. E., Lerch, M. L. F., Perevertaylo, V. L., Tome, W. A., & Rosenfeld, A. B. (2010). In vivo real-time rectal wall dosimetry for prostate radiotherapy. Physics in Medicine and Biology, 55(13), 3859-3871. https://doi.org/10.1088/0031-9155/55/13/019

In vivo real-time rectal wall dosimetry for prostate radiotherapy. / Hardcastle, Nicholas; Cutajar, Dean L.; Metcalfe, Peter E.; Lerch, Michael L F; Perevertaylo, Vladimir L.; Tome, Wolfgang A.; Rosenfeld, Anatoly B.

In: Physics in Medicine and Biology, Vol. 55, No. 13, 07.07.2010, p. 3859-3871.

Research output: Contribution to journalArticle

Hardcastle, N, Cutajar, DL, Metcalfe, PE, Lerch, MLF, Perevertaylo, VL, Tome, WA & Rosenfeld, AB 2010, 'In vivo real-time rectal wall dosimetry for prostate radiotherapy', Physics in Medicine and Biology, vol. 55, no. 13, pp. 3859-3871. https://doi.org/10.1088/0031-9155/55/13/019
Hardcastle N, Cutajar DL, Metcalfe PE, Lerch MLF, Perevertaylo VL, Tome WA et al. In vivo real-time rectal wall dosimetry for prostate radiotherapy. Physics in Medicine and Biology. 2010 Jul 7;55(13):3859-3871. https://doi.org/10.1088/0031-9155/55/13/019
Hardcastle, Nicholas ; Cutajar, Dean L. ; Metcalfe, Peter E. ; Lerch, Michael L F ; Perevertaylo, Vladimir L. ; Tome, Wolfgang A. ; Rosenfeld, Anatoly B. / In vivo real-time rectal wall dosimetry for prostate radiotherapy. In: Physics in Medicine and Biology. 2010 ; Vol. 55, No. 13. pp. 3859-3871.
@article{3a074861670141b78a0722051b553708,
title = "In vivo real-time rectal wall dosimetry for prostate radiotherapy",
abstract = "Rectal balloons are used in external beam prostate radiotherapy to provide reproducible anatomy and rectal dose reductions. This is an investigation into the combination of a MOSFET radiation detector with a rectal balloon for real-time in vivo rectal wall dosimetry. The MOSFET used in the study is a radiation detector that provides a water equivalent depth of measurement of 70 m. Two MOSFETs were combined in a face-to-face orientation. The reproducibility, sensitivity and angular dependence were measured for the dual MOSFET in a 6 MV photon beam. The dual MOSFET was combined with a rectal balloon and irradiated with hypothetical prostate treatments in a phantom. The anterior rectal wall dose was measured in real time and compared with the planning system calculated dose. The dual MOSFET showed angular dependence within ±2.5{\%} in the azimuth and +2.5{\%}/-4{\%} in the polar axes. When compared with an ion chamber measurement in a phantom, the dual MOSFET agreed within 2.5{\%} for a range of radiation path lengths and incident angles. The dual MOSFET had reproducible sensitivity for fraction sizes of 2-10 Gy. For the hypothetical prostate treatments the measured anterior rectal wall dose was 2.6 and 3.2{\%} lower than the calculated dose for 3DCRT and IMRT plans. This was expected due to limitations of the dose calculation method used at the balloon cavity interface. A dual MOSFET combined with a commercial rectal balloon was shown to provide reproducible measurements of the anterior rectal wall dose in real time. The measured anterior rectal wall dose agreed with the expected dose from the treatment plan for 3DCRT and IMRT plans. The dual MOSFET could be read out in real time during the irradiation, providing the capability for real-time dose monitoring of the rectal wall dose during treatment.",
author = "Nicholas Hardcastle and Cutajar, {Dean L.} and Metcalfe, {Peter E.} and Lerch, {Michael L F} and Perevertaylo, {Vladimir L.} and Tome, {Wolfgang A.} and Rosenfeld, {Anatoly B.}",
year = "2010",
month = "7",
day = "7",
doi = "10.1088/0031-9155/55/13/019",
language = "English (US)",
volume = "55",
pages = "3859--3871",
journal = "Physics in Medicine and Biology",
issn = "0031-9155",
publisher = "IOP Publishing Ltd.",
number = "13",

}

TY - JOUR

T1 - In vivo real-time rectal wall dosimetry for prostate radiotherapy

AU - Hardcastle, Nicholas

AU - Cutajar, Dean L.

AU - Metcalfe, Peter E.

AU - Lerch, Michael L F

AU - Perevertaylo, Vladimir L.

AU - Tome, Wolfgang A.

AU - Rosenfeld, Anatoly B.

PY - 2010/7/7

Y1 - 2010/7/7

N2 - Rectal balloons are used in external beam prostate radiotherapy to provide reproducible anatomy and rectal dose reductions. This is an investigation into the combination of a MOSFET radiation detector with a rectal balloon for real-time in vivo rectal wall dosimetry. The MOSFET used in the study is a radiation detector that provides a water equivalent depth of measurement of 70 m. Two MOSFETs were combined in a face-to-face orientation. The reproducibility, sensitivity and angular dependence were measured for the dual MOSFET in a 6 MV photon beam. The dual MOSFET was combined with a rectal balloon and irradiated with hypothetical prostate treatments in a phantom. The anterior rectal wall dose was measured in real time and compared with the planning system calculated dose. The dual MOSFET showed angular dependence within ±2.5% in the azimuth and +2.5%/-4% in the polar axes. When compared with an ion chamber measurement in a phantom, the dual MOSFET agreed within 2.5% for a range of radiation path lengths and incident angles. The dual MOSFET had reproducible sensitivity for fraction sizes of 2-10 Gy. For the hypothetical prostate treatments the measured anterior rectal wall dose was 2.6 and 3.2% lower than the calculated dose for 3DCRT and IMRT plans. This was expected due to limitations of the dose calculation method used at the balloon cavity interface. A dual MOSFET combined with a commercial rectal balloon was shown to provide reproducible measurements of the anterior rectal wall dose in real time. The measured anterior rectal wall dose agreed with the expected dose from the treatment plan for 3DCRT and IMRT plans. The dual MOSFET could be read out in real time during the irradiation, providing the capability for real-time dose monitoring of the rectal wall dose during treatment.

AB - Rectal balloons are used in external beam prostate radiotherapy to provide reproducible anatomy and rectal dose reductions. This is an investigation into the combination of a MOSFET radiation detector with a rectal balloon for real-time in vivo rectal wall dosimetry. The MOSFET used in the study is a radiation detector that provides a water equivalent depth of measurement of 70 m. Two MOSFETs were combined in a face-to-face orientation. The reproducibility, sensitivity and angular dependence were measured for the dual MOSFET in a 6 MV photon beam. The dual MOSFET was combined with a rectal balloon and irradiated with hypothetical prostate treatments in a phantom. The anterior rectal wall dose was measured in real time and compared with the planning system calculated dose. The dual MOSFET showed angular dependence within ±2.5% in the azimuth and +2.5%/-4% in the polar axes. When compared with an ion chamber measurement in a phantom, the dual MOSFET agreed within 2.5% for a range of radiation path lengths and incident angles. The dual MOSFET had reproducible sensitivity for fraction sizes of 2-10 Gy. For the hypothetical prostate treatments the measured anterior rectal wall dose was 2.6 and 3.2% lower than the calculated dose for 3DCRT and IMRT plans. This was expected due to limitations of the dose calculation method used at the balloon cavity interface. A dual MOSFET combined with a commercial rectal balloon was shown to provide reproducible measurements of the anterior rectal wall dose in real time. The measured anterior rectal wall dose agreed with the expected dose from the treatment plan for 3DCRT and IMRT plans. The dual MOSFET could be read out in real time during the irradiation, providing the capability for real-time dose monitoring of the rectal wall dose during treatment.

UR - http://www.scopus.com/inward/record.url?scp=77957935151&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=77957935151&partnerID=8YFLogxK

U2 - 10.1088/0031-9155/55/13/019

DO - 10.1088/0031-9155/55/13/019

M3 - Article

C2 - 20571209

AN - SCOPUS:77957935151

VL - 55

SP - 3859

EP - 3871

JO - Physics in Medicine and Biology

JF - Physics in Medicine and Biology

SN - 0031-9155

IS - 13

ER -