A spiral phantom for IMRT and tomotherapy treatment delivery verification

B. Paliwal, Wolfgang A. Tome, S. Richardson, T. R. Mackie

Research output: Contribution to journalArticle

20 Citations (Scopus)

Abstract

A solid water/cylindrical phantom is machined to create a spiral cavity for placing radiographic or radiochromatic film in a spiral configuration. This spiral phantom is used to sample, predict, and measure data in three-dimension subspace. The predicted data are obtained by projecting the patient plan data on the spiral phantom in the treatment planning software. The measured data are obtained by irradiating the spiral phantom (with film in the spiral cavity) as per the treatment plan. The predicted and measured data are converted to a two-dimensional matrix and plotted as a spiralogram. Comparison of these predicted and measured spiralograms provides a quantitative comparison and thus validation of treatment delivered as planned. The spiral phantom is a simple, cost-effective approach to sample 3D data from complexly shaped, intensity modulated or compensated multiple beams. A software script is being written to automate the entire process of projection, data sampling, and comparison. Design aspects and some examples of dose verification are presented. The usefulness of the spiral phantom for intensity modulated radiation therapy and dynamic field shaping are discussed. (C) 2000 American Association of Physicists in Medicine.

Original languageEnglish (US)
Pages (from-to)2503-2507
Number of pages5
JournalMedical Physics
Volume27
Issue number11
DOIs
StatePublished - 2000
Externally publishedYes

Fingerprint

Software
Radiotherapy
Therapeutics
Costs and Cost Analysis
Water

Keywords

  • IMRT treatment plan verification
  • QA phantom

ASJC Scopus subject areas

  • Biophysics

Cite this

A spiral phantom for IMRT and tomotherapy treatment delivery verification. / Paliwal, B.; Tome, Wolfgang A.; Richardson, S.; Mackie, T. R.

In: Medical Physics, Vol. 27, No. 11, 2000, p. 2503-2507.

Research output: Contribution to journalArticle

Paliwal, B. ; Tome, Wolfgang A. ; Richardson, S. ; Mackie, T. R. / A spiral phantom for IMRT and tomotherapy treatment delivery verification. In: Medical Physics. 2000 ; Vol. 27, No. 11. pp. 2503-2507.
@article{a8cecbd5ec9448518e25da144e81edab,
title = "A spiral phantom for IMRT and tomotherapy treatment delivery verification",
abstract = "A solid water/cylindrical phantom is machined to create a spiral cavity for placing radiographic or radiochromatic film in a spiral configuration. This spiral phantom is used to sample, predict, and measure data in three-dimension subspace. The predicted data are obtained by projecting the patient plan data on the spiral phantom in the treatment planning software. The measured data are obtained by irradiating the spiral phantom (with film in the spiral cavity) as per the treatment plan. The predicted and measured data are converted to a two-dimensional matrix and plotted as a spiralogram. Comparison of these predicted and measured spiralograms provides a quantitative comparison and thus validation of treatment delivered as planned. The spiral phantom is a simple, cost-effective approach to sample 3D data from complexly shaped, intensity modulated or compensated multiple beams. A software script is being written to automate the entire process of projection, data sampling, and comparison. Design aspects and some examples of dose verification are presented. The usefulness of the spiral phantom for intensity modulated radiation therapy and dynamic field shaping are discussed. (C) 2000 American Association of Physicists in Medicine.",
keywords = "IMRT treatment plan verification, QA phantom",
author = "B. Paliwal and Tome, {Wolfgang A.} and S. Richardson and Mackie, {T. R.}",
year = "2000",
doi = "10.1118/1.1319523",
language = "English (US)",
volume = "27",
pages = "2503--2507",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "11",

}

TY - JOUR

T1 - A spiral phantom for IMRT and tomotherapy treatment delivery verification

AU - Paliwal, B.

AU - Tome, Wolfgang A.

AU - Richardson, S.

AU - Mackie, T. R.

PY - 2000

Y1 - 2000

N2 - A solid water/cylindrical phantom is machined to create a spiral cavity for placing radiographic or radiochromatic film in a spiral configuration. This spiral phantom is used to sample, predict, and measure data in three-dimension subspace. The predicted data are obtained by projecting the patient plan data on the spiral phantom in the treatment planning software. The measured data are obtained by irradiating the spiral phantom (with film in the spiral cavity) as per the treatment plan. The predicted and measured data are converted to a two-dimensional matrix and plotted as a spiralogram. Comparison of these predicted and measured spiralograms provides a quantitative comparison and thus validation of treatment delivered as planned. The spiral phantom is a simple, cost-effective approach to sample 3D data from complexly shaped, intensity modulated or compensated multiple beams. A software script is being written to automate the entire process of projection, data sampling, and comparison. Design aspects and some examples of dose verification are presented. The usefulness of the spiral phantom for intensity modulated radiation therapy and dynamic field shaping are discussed. (C) 2000 American Association of Physicists in Medicine.

AB - A solid water/cylindrical phantom is machined to create a spiral cavity for placing radiographic or radiochromatic film in a spiral configuration. This spiral phantom is used to sample, predict, and measure data in three-dimension subspace. The predicted data are obtained by projecting the patient plan data on the spiral phantom in the treatment planning software. The measured data are obtained by irradiating the spiral phantom (with film in the spiral cavity) as per the treatment plan. The predicted and measured data are converted to a two-dimensional matrix and plotted as a spiralogram. Comparison of these predicted and measured spiralograms provides a quantitative comparison and thus validation of treatment delivered as planned. The spiral phantom is a simple, cost-effective approach to sample 3D data from complexly shaped, intensity modulated or compensated multiple beams. A software script is being written to automate the entire process of projection, data sampling, and comparison. Design aspects and some examples of dose verification are presented. The usefulness of the spiral phantom for intensity modulated radiation therapy and dynamic field shaping are discussed. (C) 2000 American Association of Physicists in Medicine.

KW - IMRT treatment plan verification

KW - QA phantom

UR - http://www.scopus.com/inward/record.url?scp=0033695029&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0033695029&partnerID=8YFLogxK

U2 - 10.1118/1.1319523

DO - 10.1118/1.1319523

M3 - Article

C2 - 11128301

AN - SCOPUS:0033695029

VL - 27

SP - 2503

EP - 2507

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 11

ER -