A portable microkeratome for harvesting the human anterior corneal surface

Melvin A. Sarayba, Tulaya Tungsiripat, Paula M. Sweet, Roy S. Chuck

Research output: Contribution to journalArticle

3 Citations (Scopus)

Abstract

A portable, gas-driven turbine microkeratome device capable of harvesting the entire anterior corneal surface for lamellar transplantation on human donor globes was evaluated. The device consisted of a modified LASIK microkeratome with an enlarged suction ring, head, and blade. Vacuum was achieved by a simple hand pump. Lamellar keratectomy was performed on 5 fresh human donor globes. Lenticule dimensions were measured on days 0, 3, 6, and 9 after storage in preservation media at 4°C. On day 0, the obtained lenticules were 13.9 ± 0.9 mm and 13.5 ± 0.4 mm, vertical and horizontal diameters, respectively. The average central lenticule thickness was 152.2 ± 52 μm. Each lenticule was uniform in thickness over 5 measurement points (P = 0.74). Repeat measurements of corneal thickness over the 9 days showed no statistically significant difference (P = 0.51). On day 9 lenticules were 14.6 ± 0.3 mm and 14.6 ± 0.4 mm, vertical and horizontal diameters, respectively. When day 0 was compared to day 9, vertical diameter also showed no statistically significant difference (P = 0.16), whereas horizontal diameter was significantly different (P < 0.001). This device proves to be an economical alternative to electric-powered systems for the harvest of transplantable corneal sections.

Original languageEnglish (US)
Pages (from-to)443-446
Number of pages4
JournalCornea
Volume23
Issue number5
DOIs
StatePublished - Jul 2004
Externally publishedYes

Fingerprint

Equipment and Supplies
Corneal Pachymetry
Laser In Situ Keratomileusis
Suction
Vacuum
Hand
Transplantation
Gases
Head

Keywords

  • Anterior corneal surface
  • Donor corneas
  • Microkeratome

ASJC Scopus subject areas

  • Ophthalmology

Cite this

A portable microkeratome for harvesting the human anterior corneal surface. / Sarayba, Melvin A.; Tungsiripat, Tulaya; Sweet, Paula M.; Chuck, Roy S.

In: Cornea, Vol. 23, No. 5, 07.2004, p. 443-446.

Research output: Contribution to journalArticle

Sarayba, Melvin A. ; Tungsiripat, Tulaya ; Sweet, Paula M. ; Chuck, Roy S. / A portable microkeratome for harvesting the human anterior corneal surface. In: Cornea. 2004 ; Vol. 23, No. 5. pp. 443-446.
@article{cf131ae83e244765bf5924cd183f6b40,
title = "A portable microkeratome for harvesting the human anterior corneal surface",
abstract = "A portable, gas-driven turbine microkeratome device capable of harvesting the entire anterior corneal surface for lamellar transplantation on human donor globes was evaluated. The device consisted of a modified LASIK microkeratome with an enlarged suction ring, head, and blade. Vacuum was achieved by a simple hand pump. Lamellar keratectomy was performed on 5 fresh human donor globes. Lenticule dimensions were measured on days 0, 3, 6, and 9 after storage in preservation media at 4°C. On day 0, the obtained lenticules were 13.9 ± 0.9 mm and 13.5 ± 0.4 mm, vertical and horizontal diameters, respectively. The average central lenticule thickness was 152.2 ± 52 μm. Each lenticule was uniform in thickness over 5 measurement points (P = 0.74). Repeat measurements of corneal thickness over the 9 days showed no statistically significant difference (P = 0.51). On day 9 lenticules were 14.6 ± 0.3 mm and 14.6 ± 0.4 mm, vertical and horizontal diameters, respectively. When day 0 was compared to day 9, vertical diameter also showed no statistically significant difference (P = 0.16), whereas horizontal diameter was significantly different (P < 0.001). This device proves to be an economical alternative to electric-powered systems for the harvest of transplantable corneal sections.",
keywords = "Anterior corneal surface, Donor corneas, Microkeratome",
author = "Sarayba, {Melvin A.} and Tulaya Tungsiripat and Sweet, {Paula M.} and Chuck, {Roy S.}",
year = "2004",
month = "7",
doi = "10.1097/01.ico.0000122701.49054.c9",
language = "English (US)",
volume = "23",
pages = "443--446",
journal = "Cornea",
issn = "0277-3740",
publisher = "Lippincott Williams and Wilkins",
number = "5",

}

TY - JOUR

T1 - A portable microkeratome for harvesting the human anterior corneal surface

AU - Sarayba, Melvin A.

AU - Tungsiripat, Tulaya

AU - Sweet, Paula M.

AU - Chuck, Roy S.

PY - 2004/7

Y1 - 2004/7

N2 - A portable, gas-driven turbine microkeratome device capable of harvesting the entire anterior corneal surface for lamellar transplantation on human donor globes was evaluated. The device consisted of a modified LASIK microkeratome with an enlarged suction ring, head, and blade. Vacuum was achieved by a simple hand pump. Lamellar keratectomy was performed on 5 fresh human donor globes. Lenticule dimensions were measured on days 0, 3, 6, and 9 after storage in preservation media at 4°C. On day 0, the obtained lenticules were 13.9 ± 0.9 mm and 13.5 ± 0.4 mm, vertical and horizontal diameters, respectively. The average central lenticule thickness was 152.2 ± 52 μm. Each lenticule was uniform in thickness over 5 measurement points (P = 0.74). Repeat measurements of corneal thickness over the 9 days showed no statistically significant difference (P = 0.51). On day 9 lenticules were 14.6 ± 0.3 mm and 14.6 ± 0.4 mm, vertical and horizontal diameters, respectively. When day 0 was compared to day 9, vertical diameter also showed no statistically significant difference (P = 0.16), whereas horizontal diameter was significantly different (P < 0.001). This device proves to be an economical alternative to electric-powered systems for the harvest of transplantable corneal sections.

AB - A portable, gas-driven turbine microkeratome device capable of harvesting the entire anterior corneal surface for lamellar transplantation on human donor globes was evaluated. The device consisted of a modified LASIK microkeratome with an enlarged suction ring, head, and blade. Vacuum was achieved by a simple hand pump. Lamellar keratectomy was performed on 5 fresh human donor globes. Lenticule dimensions were measured on days 0, 3, 6, and 9 after storage in preservation media at 4°C. On day 0, the obtained lenticules were 13.9 ± 0.9 mm and 13.5 ± 0.4 mm, vertical and horizontal diameters, respectively. The average central lenticule thickness was 152.2 ± 52 μm. Each lenticule was uniform in thickness over 5 measurement points (P = 0.74). Repeat measurements of corneal thickness over the 9 days showed no statistically significant difference (P = 0.51). On day 9 lenticules were 14.6 ± 0.3 mm and 14.6 ± 0.4 mm, vertical and horizontal diameters, respectively. When day 0 was compared to day 9, vertical diameter also showed no statistically significant difference (P = 0.16), whereas horizontal diameter was significantly different (P < 0.001). This device proves to be an economical alternative to electric-powered systems for the harvest of transplantable corneal sections.

KW - Anterior corneal surface

KW - Donor corneas

KW - Microkeratome

UR - http://www.scopus.com/inward/record.url?scp=3042784589&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=3042784589&partnerID=8YFLogxK

U2 - 10.1097/01.ico.0000122701.49054.c9

DO - 10.1097/01.ico.0000122701.49054.c9

M3 - Article

C2 - 15220727

AN - SCOPUS:3042784589

VL - 23

SP - 443

EP - 446

JO - Cornea

JF - Cornea

SN - 0277-3740

IS - 5

ER -