TU‐G‐BRB‐06: Conversion of X‐Ray Sinogram to Proton Sinogram to Address the Limited‐Range Issue in Proton Computed Tomography

D. Wang, T. Mackie, Wolfgang A. Tome

Research output: Contribution to journalArticle

Abstract

Purpose: To investigate a novel methodology of converting x‐ray sinograms into proton sinogram, which can be used to replace the missing part of an actual proton sinogram due to the limited‐range problem. Methods: Modeling has shown that there exist correlations between proton projection and multispectral x‐ray projection for materials in the human body. Based on these correlations, materials can be categorized into: A) adipose tissue, B) non‐adipose soft tissue, C) bone. These three categories can be identified on a regular x‐ray computed tomography (xCT) image without much effort. With an xCT image and its sinogram available, the x‐ray sinogram can be converted to the corresponding proton sinogram by: 1) converting the x‐ray raysum through adipose and non‐adipose soft tissues using the predetermined relationships and the relative portions of two types of material along the ray; and 2) by correcting for bone if the ray passes through it. In a proof‐of‐principle study, numerical simulations were done. The converted proton sinogram was compared to the simulated one. The converted proton sinogram is used to make up the 200 MeV proton projections through water‐equivalent depth (WED) larger than 20 cm, which are then combined with simulated proton projections through WED < 20 cm for pCT stopping power reconstruction. Results: A sinogram profile for view angle = 0 showed the converted projections had a mean error of −0.13% (standard deviation 1.7%), RMS error of 0.20%, compared to the simulated one. The pCT stopping power image reconstructed using the sinogram with converted part (13.5% of total projections) showed a mean error of only +0.3% (standard deviation 4.6%). Conclusions: The proposed methodology allows one to convert x‐ray sinograms into proton sinograms with a reasonable accuracy, and is particularly useful to address the limited‐range problem of pCT scanning.

Original languageEnglish (US)
Pages (from-to)3779-3780
Number of pages2
JournalMedical Physics
Volume38
Issue number6
DOIs
StatePublished - 2011
Externally publishedYes

Fingerprint

Protons
Tomography
X-Rays
X Ray Computed Tomography
Bone and Bones
Human Body
Adipose Tissue

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

TU‐G‐BRB‐06 : Conversion of X‐Ray Sinogram to Proton Sinogram to Address the Limited‐Range Issue in Proton Computed Tomography. / Wang, D.; Mackie, T.; Tome, Wolfgang A.

In: Medical Physics, Vol. 38, No. 6, 2011, p. 3779-3780.

Research output: Contribution to journalArticle

@article{f314fe01dd364f50ad1307b41b3726b8,
title = "TU‐G‐BRB‐06: Conversion of X‐Ray Sinogram to Proton Sinogram to Address the Limited‐Range Issue in Proton Computed Tomography",
abstract = "Purpose: To investigate a novel methodology of converting x‐ray sinograms into proton sinogram, which can be used to replace the missing part of an actual proton sinogram due to the limited‐range problem. Methods: Modeling has shown that there exist correlations between proton projection and multispectral x‐ray projection for materials in the human body. Based on these correlations, materials can be categorized into: A) adipose tissue, B) non‐adipose soft tissue, C) bone. These three categories can be identified on a regular x‐ray computed tomography (xCT) image without much effort. With an xCT image and its sinogram available, the x‐ray sinogram can be converted to the corresponding proton sinogram by: 1) converting the x‐ray raysum through adipose and non‐adipose soft tissues using the predetermined relationships and the relative portions of two types of material along the ray; and 2) by correcting for bone if the ray passes through it. In a proof‐of‐principle study, numerical simulations were done. The converted proton sinogram was compared to the simulated one. The converted proton sinogram is used to make up the 200 MeV proton projections through water‐equivalent depth (WED) larger than 20 cm, which are then combined with simulated proton projections through WED < 20 cm for pCT stopping power reconstruction. Results: A sinogram profile for view angle = 0 showed the converted projections had a mean error of −0.13{\%} (standard deviation 1.7{\%}), RMS error of 0.20{\%}, compared to the simulated one. The pCT stopping power image reconstructed using the sinogram with converted part (13.5{\%} of total projections) showed a mean error of only +0.3{\%} (standard deviation 4.6{\%}). Conclusions: The proposed methodology allows one to convert x‐ray sinograms into proton sinograms with a reasonable accuracy, and is particularly useful to address the limited‐range problem of pCT scanning.",
author = "D. Wang and T. Mackie and Tome, {Wolfgang A.}",
year = "2011",
doi = "10.1118/1.3613226",
language = "English (US)",
volume = "38",
pages = "3779--3780",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "6",

}

TY - JOUR

T1 - TU‐G‐BRB‐06

T2 - Conversion of X‐Ray Sinogram to Proton Sinogram to Address the Limited‐Range Issue in Proton Computed Tomography

AU - Wang, D.

AU - Mackie, T.

AU - Tome, Wolfgang A.

PY - 2011

Y1 - 2011

N2 - Purpose: To investigate a novel methodology of converting x‐ray sinograms into proton sinogram, which can be used to replace the missing part of an actual proton sinogram due to the limited‐range problem. Methods: Modeling has shown that there exist correlations between proton projection and multispectral x‐ray projection for materials in the human body. Based on these correlations, materials can be categorized into: A) adipose tissue, B) non‐adipose soft tissue, C) bone. These three categories can be identified on a regular x‐ray computed tomography (xCT) image without much effort. With an xCT image and its sinogram available, the x‐ray sinogram can be converted to the corresponding proton sinogram by: 1) converting the x‐ray raysum through adipose and non‐adipose soft tissues using the predetermined relationships and the relative portions of two types of material along the ray; and 2) by correcting for bone if the ray passes through it. In a proof‐of‐principle study, numerical simulations were done. The converted proton sinogram was compared to the simulated one. The converted proton sinogram is used to make up the 200 MeV proton projections through water‐equivalent depth (WED) larger than 20 cm, which are then combined with simulated proton projections through WED < 20 cm for pCT stopping power reconstruction. Results: A sinogram profile for view angle = 0 showed the converted projections had a mean error of −0.13% (standard deviation 1.7%), RMS error of 0.20%, compared to the simulated one. The pCT stopping power image reconstructed using the sinogram with converted part (13.5% of total projections) showed a mean error of only +0.3% (standard deviation 4.6%). Conclusions: The proposed methodology allows one to convert x‐ray sinograms into proton sinograms with a reasonable accuracy, and is particularly useful to address the limited‐range problem of pCT scanning.

AB - Purpose: To investigate a novel methodology of converting x‐ray sinograms into proton sinogram, which can be used to replace the missing part of an actual proton sinogram due to the limited‐range problem. Methods: Modeling has shown that there exist correlations between proton projection and multispectral x‐ray projection for materials in the human body. Based on these correlations, materials can be categorized into: A) adipose tissue, B) non‐adipose soft tissue, C) bone. These three categories can be identified on a regular x‐ray computed tomography (xCT) image without much effort. With an xCT image and its sinogram available, the x‐ray sinogram can be converted to the corresponding proton sinogram by: 1) converting the x‐ray raysum through adipose and non‐adipose soft tissues using the predetermined relationships and the relative portions of two types of material along the ray; and 2) by correcting for bone if the ray passes through it. In a proof‐of‐principle study, numerical simulations were done. The converted proton sinogram was compared to the simulated one. The converted proton sinogram is used to make up the 200 MeV proton projections through water‐equivalent depth (WED) larger than 20 cm, which are then combined with simulated proton projections through WED < 20 cm for pCT stopping power reconstruction. Results: A sinogram profile for view angle = 0 showed the converted projections had a mean error of −0.13% (standard deviation 1.7%), RMS error of 0.20%, compared to the simulated one. The pCT stopping power image reconstructed using the sinogram with converted part (13.5% of total projections) showed a mean error of only +0.3% (standard deviation 4.6%). Conclusions: The proposed methodology allows one to convert x‐ray sinograms into proton sinograms with a reasonable accuracy, and is particularly useful to address the limited‐range problem of pCT scanning.

UR - http://www.scopus.com/inward/record.url?scp=85024813827&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85024813827&partnerID=8YFLogxK

U2 - 10.1118/1.3613226

DO - 10.1118/1.3613226

M3 - Article

AN - SCOPUS:85024813827

VL - 38

SP - 3779

EP - 3780

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 6

ER -