Transvenous dual site left ventricular pacing to lower elevated pacing thresholds

Melissa E. Harding, Eugen C. Palma

Research output: Contribution to journalArticle

Abstract

Dual site left ventricular pacing through two left ventricular pacing leads, located in discrete vessels, significantly lowered pacing thresholds from 6 V at 1 ms and 4.25V at 0.5 ms through the leads individually, to 0.75V at 0.5 ms by utilizing a Y-adaptor to connect the two leads. (PACE 2011; e6-e8)

Original languageEnglish (US)
JournalPACE - Pacing and Clinical Electrophysiology
Volume34
Issue number1
DOIs
StatePublished - Jan 2011

Keywords

  • CRT
  • pacing

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine

Cite this

Transvenous dual site left ventricular pacing to lower elevated pacing thresholds. / Harding, Melissa E.; Palma, Eugen C.

In: PACE - Pacing and Clinical Electrophysiology, Vol. 34, No. 1, 01.2011.

Research output: Contribution to journalArticle

@article{b672b91ba647495eb7f99fdf038ef82a,
title = "Transvenous dual site left ventricular pacing to lower elevated pacing thresholds",
abstract = "Dual site left ventricular pacing through two left ventricular pacing leads, located in discrete vessels, significantly lowered pacing thresholds from 6 V at 1 ms and 4.25V at 0.5 ms through the leads individually, to 0.75V at 0.5 ms by utilizing a Y-adaptor to connect the two leads. (PACE 2011; e6-e8)",
keywords = "CRT, pacing",
author = "Harding, {Melissa E.} and Palma, {Eugen C.}",
year = "2011",
month = "1",
doi = "10.1111/j.1540-8159.2010.02736.x",
language = "English (US)",
volume = "34",
journal = "PACE - Pacing and Clinical Electrophysiology",
issn = "0147-8389",
publisher = "Wiley-Blackwell",
number = "1",

}

TY - JOUR

T1 - Transvenous dual site left ventricular pacing to lower elevated pacing thresholds

AU - Harding, Melissa E.

AU - Palma, Eugen C.

PY - 2011/1

Y1 - 2011/1

N2 - Dual site left ventricular pacing through two left ventricular pacing leads, located in discrete vessels, significantly lowered pacing thresholds from 6 V at 1 ms and 4.25V at 0.5 ms through the leads individually, to 0.75V at 0.5 ms by utilizing a Y-adaptor to connect the two leads. (PACE 2011; e6-e8)

AB - Dual site left ventricular pacing through two left ventricular pacing leads, located in discrete vessels, significantly lowered pacing thresholds from 6 V at 1 ms and 4.25V at 0.5 ms through the leads individually, to 0.75V at 0.5 ms by utilizing a Y-adaptor to connect the two leads. (PACE 2011; e6-e8)

KW - CRT

KW - pacing

UR - http://www.scopus.com/inward/record.url?scp=78651328497&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=78651328497&partnerID=8YFLogxK

U2 - 10.1111/j.1540-8159.2010.02736.x

DO - 10.1111/j.1540-8159.2010.02736.x

M3 - Article

VL - 34

JO - PACE - Pacing and Clinical Electrophysiology

JF - PACE - Pacing and Clinical Electrophysiology

SN - 0147-8389

IS - 1

ER -