The use of diode dosimetry in quality improvement of patient care in radiation therapy

Doracy P. Fontenla, Ravindra Yaparpalvi, Chen Shou Chui, Edith Briot

Research output: Contribution to journalArticle

13 Citations (Scopus)

Abstract

The purpose of this work is to improve the quality of patient care in radiation therapy by implementing a comprehensive quality assurance (QA) program aiming to enhance patient in vivo dosimetry on a routine basis. The characteristics of two commercially available semi-conductor diode dosimetry systems were evaluated. The diodes were calibrated relative to an ionization chamber-electrometer system with calibrations traceable to the National Institute of Standards and Technology (NIST). Correction factors of clinical relevance were quantified to convert the diode readings into patient dose. The results of dose measurements on 6 patients undergoing external beam radiation therapy for carcinoma of the prostate on three different therapy units are presented. Field shaping during treatments was accomplished either by multileaf collimation or by cerrobend blocking. A deviation of less than ±4% between the measured and prescribed patient doses was observed. The results indicate that the diodes exhibit excellent linearity, dose reproducibility, minimal anisotropy, and can be used with confidence for patient dose verification. Furthermore, diodes render real time verification of dose delivered to patients.

Original languageEnglish (US)
Pages (from-to)235-241
Number of pages7
JournalMedical Dosimetry
Volume21
Issue number4
DOIs
StatePublished - Dec 1996
Externally publishedYes

Fingerprint

Quality Improvement
dosimeters
radiation therapy
Patient Care
Radiotherapy
diodes
dosage
Quality of Health Care
Anisotropy
electrometers
semiconductor diodes
assurance
collimation
Calibration
ionization chambers
Reading
Prostate
linearity
confidence
therapy

Keywords

  • Diode dosimetry
  • External beam radiation therapy
  • In-vivo dosimetry
  • Prostate carcinoma
  • QA of patient care

ASJC Scopus subject areas

  • Oncology
  • Radiology Nuclear Medicine and imaging
  • Radiation
  • Health Professions(all)

Cite this

The use of diode dosimetry in quality improvement of patient care in radiation therapy. / Fontenla, Doracy P.; Yaparpalvi, Ravindra; Chui, Chen Shou; Briot, Edith.

In: Medical Dosimetry, Vol. 21, No. 4, 12.1996, p. 235-241.

Research output: Contribution to journalArticle

Fontenla, Doracy P. ; Yaparpalvi, Ravindra ; Chui, Chen Shou ; Briot, Edith. / The use of diode dosimetry in quality improvement of patient care in radiation therapy. In: Medical Dosimetry. 1996 ; Vol. 21, No. 4. pp. 235-241.
@article{cf7a77dc16594f65b35b843fad7966d7,
title = "The use of diode dosimetry in quality improvement of patient care in radiation therapy",
abstract = "The purpose of this work is to improve the quality of patient care in radiation therapy by implementing a comprehensive quality assurance (QA) program aiming to enhance patient in vivo dosimetry on a routine basis. The characteristics of two commercially available semi-conductor diode dosimetry systems were evaluated. The diodes were calibrated relative to an ionization chamber-electrometer system with calibrations traceable to the National Institute of Standards and Technology (NIST). Correction factors of clinical relevance were quantified to convert the diode readings into patient dose. The results of dose measurements on 6 patients undergoing external beam radiation therapy for carcinoma of the prostate on three different therapy units are presented. Field shaping during treatments was accomplished either by multileaf collimation or by cerrobend blocking. A deviation of less than ±4{\%} between the measured and prescribed patient doses was observed. The results indicate that the diodes exhibit excellent linearity, dose reproducibility, minimal anisotropy, and can be used with confidence for patient dose verification. Furthermore, diodes render real time verification of dose delivered to patients.",
keywords = "Diode dosimetry, External beam radiation therapy, In-vivo dosimetry, Prostate carcinoma, QA of patient care",
author = "Fontenla, {Doracy P.} and Ravindra Yaparpalvi and Chui, {Chen Shou} and Edith Briot",
year = "1996",
month = "12",
doi = "10.1016/S0958-3947(96)00081-7",
language = "English (US)",
volume = "21",
pages = "235--241",
journal = "Medical Dosimetry",
issn = "0958-3947",
publisher = "Elsevier Inc.",
number = "4",

}

TY - JOUR

T1 - The use of diode dosimetry in quality improvement of patient care in radiation therapy

AU - Fontenla, Doracy P.

AU - Yaparpalvi, Ravindra

AU - Chui, Chen Shou

AU - Briot, Edith

PY - 1996/12

Y1 - 1996/12

N2 - The purpose of this work is to improve the quality of patient care in radiation therapy by implementing a comprehensive quality assurance (QA) program aiming to enhance patient in vivo dosimetry on a routine basis. The characteristics of two commercially available semi-conductor diode dosimetry systems were evaluated. The diodes were calibrated relative to an ionization chamber-electrometer system with calibrations traceable to the National Institute of Standards and Technology (NIST). Correction factors of clinical relevance were quantified to convert the diode readings into patient dose. The results of dose measurements on 6 patients undergoing external beam radiation therapy for carcinoma of the prostate on three different therapy units are presented. Field shaping during treatments was accomplished either by multileaf collimation or by cerrobend blocking. A deviation of less than ±4% between the measured and prescribed patient doses was observed. The results indicate that the diodes exhibit excellent linearity, dose reproducibility, minimal anisotropy, and can be used with confidence for patient dose verification. Furthermore, diodes render real time verification of dose delivered to patients.

AB - The purpose of this work is to improve the quality of patient care in radiation therapy by implementing a comprehensive quality assurance (QA) program aiming to enhance patient in vivo dosimetry on a routine basis. The characteristics of two commercially available semi-conductor diode dosimetry systems were evaluated. The diodes were calibrated relative to an ionization chamber-electrometer system with calibrations traceable to the National Institute of Standards and Technology (NIST). Correction factors of clinical relevance were quantified to convert the diode readings into patient dose. The results of dose measurements on 6 patients undergoing external beam radiation therapy for carcinoma of the prostate on three different therapy units are presented. Field shaping during treatments was accomplished either by multileaf collimation or by cerrobend blocking. A deviation of less than ±4% between the measured and prescribed patient doses was observed. The results indicate that the diodes exhibit excellent linearity, dose reproducibility, minimal anisotropy, and can be used with confidence for patient dose verification. Furthermore, diodes render real time verification of dose delivered to patients.

KW - Diode dosimetry

KW - External beam radiation therapy

KW - In-vivo dosimetry

KW - Prostate carcinoma

KW - QA of patient care

UR - http://www.scopus.com/inward/record.url?scp=0030332980&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0030332980&partnerID=8YFLogxK

U2 - 10.1016/S0958-3947(96)00081-7

DO - 10.1016/S0958-3947(96)00081-7

M3 - Article

C2 - 8985929

AN - SCOPUS:0030332980

VL - 21

SP - 235

EP - 241

JO - Medical Dosimetry

JF - Medical Dosimetry

SN - 0958-3947

IS - 4

ER -