SU‐E‐T‐143

Quality Assurance of Complex Rotational and Stereotactic IMRT Treatments Using Gafchromic EBT2 Film

T. Kairn, T. Aland, N. Hardcastle, B. Hill, J. Kenny, R. Knight, R. Meldrum, N. Middlebrook, Wolfgang A. Tome

Research output: Contribution to journalArticle

Abstract

Purpose: The accurate completion of reliable quality assurance measurements of complex intensity‐modulated and rotational radiotherapy treatments demands the use of a dosimetry system with both high resolution and minimal angular dependence. This study demonstrates the utility of Gafchromic EBT2 radiochromic dosimetry film for the quality assurance of complex radiotherapy treatments. Methods: This study establishes that the dose‐response of EBT2 film has minimal angular and depth dependence and then demonstrates how the film can be used to perform quality assurance measurements for BrainLab stereotactic IMRT, Varian RapidArc and TomoTherapy treatments. Results: For all treatment plans, the proportion of dose points agreeing with the film measurements to within gamma(3%,3mm) was found to be above 95%, with all points agreeing within 5%, indicating that the treatments were delivered with an acceptable level of accuracy. The film images also provided additional information on low‐level dose variations including: the location and extent of dose from inter‐leaf leakage (for RapidArc); the location and extent of helical field junctioning effects, or threading (for TomoTherapy); and the existence of small regions of under‐dosage from very small treatment segments (for stereotactic IMRT). Conclusions: The fact that these film measurements detected small‐scale dosimetric effects that were not apparent in the dose planes exported from the treatment planning systems confirms both the value of high‐resolution dosimetry and the importance of designing complex radiotherapy treatments with the maximum achievable level of target dose homogeneity.

Original languageEnglish (US)
Pages (from-to)3519
Number of pages1
JournalMedical Physics
Volume38
Issue number6
DOIs
StatePublished - 2011
Externally publishedYes

Fingerprint

Radiotherapy
Therapeutics
Film Dosimetry
Motion Pictures

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

SU‐E‐T‐143 : Quality Assurance of Complex Rotational and Stereotactic IMRT Treatments Using Gafchromic EBT2 Film. / Kairn, T.; Aland, T.; Hardcastle, N.; Hill, B.; Kenny, J.; Knight, R.; Meldrum, R.; Middlebrook, N.; Tome, Wolfgang A.

In: Medical Physics, Vol. 38, No. 6, 2011, p. 3519.

Research output: Contribution to journalArticle

Kairn, T, Aland, T, Hardcastle, N, Hill, B, Kenny, J, Knight, R, Meldrum, R, Middlebrook, N & Tome, WA 2011, 'SU‐E‐T‐143: Quality Assurance of Complex Rotational and Stereotactic IMRT Treatments Using Gafchromic EBT2 Film', Medical Physics, vol. 38, no. 6, pp. 3519. https://doi.org/10.1118/1.3612094
Kairn, T. ; Aland, T. ; Hardcastle, N. ; Hill, B. ; Kenny, J. ; Knight, R. ; Meldrum, R. ; Middlebrook, N. ; Tome, Wolfgang A. / SU‐E‐T‐143 : Quality Assurance of Complex Rotational and Stereotactic IMRT Treatments Using Gafchromic EBT2 Film. In: Medical Physics. 2011 ; Vol. 38, No. 6. pp. 3519.
@article{8a95e82238dc4c12b6dc016c1198fec6,
title = "SU‐E‐T‐143: Quality Assurance of Complex Rotational and Stereotactic IMRT Treatments Using Gafchromic EBT2 Film",
abstract = "Purpose: The accurate completion of reliable quality assurance measurements of complex intensity‐modulated and rotational radiotherapy treatments demands the use of a dosimetry system with both high resolution and minimal angular dependence. This study demonstrates the utility of Gafchromic EBT2 radiochromic dosimetry film for the quality assurance of complex radiotherapy treatments. Methods: This study establishes that the dose‐response of EBT2 film has minimal angular and depth dependence and then demonstrates how the film can be used to perform quality assurance measurements for BrainLab stereotactic IMRT, Varian RapidArc and TomoTherapy treatments. Results: For all treatment plans, the proportion of dose points agreeing with the film measurements to within gamma(3{\%},3mm) was found to be above 95{\%}, with all points agreeing within 5{\%}, indicating that the treatments were delivered with an acceptable level of accuracy. The film images also provided additional information on low‐level dose variations including: the location and extent of dose from inter‐leaf leakage (for RapidArc); the location and extent of helical field junctioning effects, or threading (for TomoTherapy); and the existence of small regions of under‐dosage from very small treatment segments (for stereotactic IMRT). Conclusions: The fact that these film measurements detected small‐scale dosimetric effects that were not apparent in the dose planes exported from the treatment planning systems confirms both the value of high‐resolution dosimetry and the importance of designing complex radiotherapy treatments with the maximum achievable level of target dose homogeneity.",
author = "T. Kairn and T. Aland and N. Hardcastle and B. Hill and J. Kenny and R. Knight and R. Meldrum and N. Middlebrook and Tome, {Wolfgang A.}",
year = "2011",
doi = "10.1118/1.3612094",
language = "English (US)",
volume = "38",
pages = "3519",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "6",

}

TY - JOUR

T1 - SU‐E‐T‐143

T2 - Quality Assurance of Complex Rotational and Stereotactic IMRT Treatments Using Gafchromic EBT2 Film

AU - Kairn, T.

AU - Aland, T.

AU - Hardcastle, N.

AU - Hill, B.

AU - Kenny, J.

AU - Knight, R.

AU - Meldrum, R.

AU - Middlebrook, N.

AU - Tome, Wolfgang A.

PY - 2011

Y1 - 2011

N2 - Purpose: The accurate completion of reliable quality assurance measurements of complex intensity‐modulated and rotational radiotherapy treatments demands the use of a dosimetry system with both high resolution and minimal angular dependence. This study demonstrates the utility of Gafchromic EBT2 radiochromic dosimetry film for the quality assurance of complex radiotherapy treatments. Methods: This study establishes that the dose‐response of EBT2 film has minimal angular and depth dependence and then demonstrates how the film can be used to perform quality assurance measurements for BrainLab stereotactic IMRT, Varian RapidArc and TomoTherapy treatments. Results: For all treatment plans, the proportion of dose points agreeing with the film measurements to within gamma(3%,3mm) was found to be above 95%, with all points agreeing within 5%, indicating that the treatments were delivered with an acceptable level of accuracy. The film images also provided additional information on low‐level dose variations including: the location and extent of dose from inter‐leaf leakage (for RapidArc); the location and extent of helical field junctioning effects, or threading (for TomoTherapy); and the existence of small regions of under‐dosage from very small treatment segments (for stereotactic IMRT). Conclusions: The fact that these film measurements detected small‐scale dosimetric effects that were not apparent in the dose planes exported from the treatment planning systems confirms both the value of high‐resolution dosimetry and the importance of designing complex radiotherapy treatments with the maximum achievable level of target dose homogeneity.

AB - Purpose: The accurate completion of reliable quality assurance measurements of complex intensity‐modulated and rotational radiotherapy treatments demands the use of a dosimetry system with both high resolution and minimal angular dependence. This study demonstrates the utility of Gafchromic EBT2 radiochromic dosimetry film for the quality assurance of complex radiotherapy treatments. Methods: This study establishes that the dose‐response of EBT2 film has minimal angular and depth dependence and then demonstrates how the film can be used to perform quality assurance measurements for BrainLab stereotactic IMRT, Varian RapidArc and TomoTherapy treatments. Results: For all treatment plans, the proportion of dose points agreeing with the film measurements to within gamma(3%,3mm) was found to be above 95%, with all points agreeing within 5%, indicating that the treatments were delivered with an acceptable level of accuracy. The film images also provided additional information on low‐level dose variations including: the location and extent of dose from inter‐leaf leakage (for RapidArc); the location and extent of helical field junctioning effects, or threading (for TomoTherapy); and the existence of small regions of under‐dosage from very small treatment segments (for stereotactic IMRT). Conclusions: The fact that these film measurements detected small‐scale dosimetric effects that were not apparent in the dose planes exported from the treatment planning systems confirms both the value of high‐resolution dosimetry and the importance of designing complex radiotherapy treatments with the maximum achievable level of target dose homogeneity.

UR - http://www.scopus.com/inward/record.url?scp=85024776788&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85024776788&partnerID=8YFLogxK

U2 - 10.1118/1.3612094

DO - 10.1118/1.3612094

M3 - Article

VL - 38

SP - 3519

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 6

ER -