Spontaneous versus extraction related injuries associated with Accufix J-wire atrial pacemaker lead: Tracking changes in patient management

D. T. Kawanishi, J. A. Brinker, R. Reeves, G. N. Kay, Jay N. Gross, G. Pioger, J. C. Petitot, A. Esler, G. Grunkemeier

Research output: Contribution to journalArticle

13 Citations (Scopus)

Abstract

To make recommendations for management of potentially fatal failure of the Accufix series of atrial J-wire permanent pacemaker leads, we closely monitored the number of injuries and fatalities resulting either from spontaneous fracture of the J-wire or from attempts to extract the lead. In a population of 30,357 patients, 2,298 patients are enrolled in a prospective follow-up Multicenter Study, the remainder are patients with known clinical status from voluntary reporting, and 2,992 patients died following implant. In the remaining 27,365 patients, 6 deaths have been attributed to J-wire related injury (J-inj) while 13 were complications (E-inj) associated with 4,076 lead extraction procedures (3,974 intravascular (intra)/102 primary thoracotomy (PT). The date of occurrences were from 1994 to November 1997. Conclusions: (1) Since lead extractions were not conducted in a controlled study, it is not known whether the deaths associated with lead extraction is in excess of what would have occurred if these leads had not been removed in this specific subset. (2) Awareness of the procedure related complication rate appears to have moderated the rate of lead extraction and may ultimately lead to management that reasonably balances the risks of patient injury.

Original languageEnglish (US)
Pages (from-to)2314-2317
Number of pages4
JournalPACE - Pacing and Clinical Electrophysiology
Volume21
Issue number11 II
StatePublished - 1998
Externally publishedYes

Fingerprint

Wounds and Injuries
Spontaneous Fractures
Thoracotomy
Multicenter Studies
Lead
Population

Keywords

  • Clinical follow-up
  • Device failure
  • Lead extraction
  • Permanent pacemaker lead

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine

Cite this

Spontaneous versus extraction related injuries associated with Accufix J-wire atrial pacemaker lead : Tracking changes in patient management. / Kawanishi, D. T.; Brinker, J. A.; Reeves, R.; Kay, G. N.; Gross, Jay N.; Pioger, G.; Petitot, J. C.; Esler, A.; Grunkemeier, G.

In: PACE - Pacing and Clinical Electrophysiology, Vol. 21, No. 11 II, 1998, p. 2314-2317.

Research output: Contribution to journalArticle

Kawanishi, DT, Brinker, JA, Reeves, R, Kay, GN, Gross, JN, Pioger, G, Petitot, JC, Esler, A & Grunkemeier, G 1998, 'Spontaneous versus extraction related injuries associated with Accufix J-wire atrial pacemaker lead: Tracking changes in patient management', PACE - Pacing and Clinical Electrophysiology, vol. 21, no. 11 II, pp. 2314-2317.
Kawanishi, D. T. ; Brinker, J. A. ; Reeves, R. ; Kay, G. N. ; Gross, Jay N. ; Pioger, G. ; Petitot, J. C. ; Esler, A. ; Grunkemeier, G. / Spontaneous versus extraction related injuries associated with Accufix J-wire atrial pacemaker lead : Tracking changes in patient management. In: PACE - Pacing and Clinical Electrophysiology. 1998 ; Vol. 21, No. 11 II. pp. 2314-2317.
@article{87574209c037445c99e50b9557cce0f1,
title = "Spontaneous versus extraction related injuries associated with Accufix J-wire atrial pacemaker lead: Tracking changes in patient management",
abstract = "To make recommendations for management of potentially fatal failure of the Accufix series of atrial J-wire permanent pacemaker leads, we closely monitored the number of injuries and fatalities resulting either from spontaneous fracture of the J-wire or from attempts to extract the lead. In a population of 30,357 patients, 2,298 patients are enrolled in a prospective follow-up Multicenter Study, the remainder are patients with known clinical status from voluntary reporting, and 2,992 patients died following implant. In the remaining 27,365 patients, 6 deaths have been attributed to J-wire related injury (J-inj) while 13 were complications (E-inj) associated with 4,076 lead extraction procedures (3,974 intravascular (intra)/102 primary thoracotomy (PT). The date of occurrences were from 1994 to November 1997. Conclusions: (1) Since lead extractions were not conducted in a controlled study, it is not known whether the deaths associated with lead extraction is in excess of what would have occurred if these leads had not been removed in this specific subset. (2) Awareness of the procedure related complication rate appears to have moderated the rate of lead extraction and may ultimately lead to management that reasonably balances the risks of patient injury.",
keywords = "Clinical follow-up, Device failure, Lead extraction, Permanent pacemaker lead",
author = "Kawanishi, {D. T.} and Brinker, {J. A.} and R. Reeves and Kay, {G. N.} and Gross, {Jay N.} and G. Pioger and Petitot, {J. C.} and A. Esler and G. Grunkemeier",
year = "1998",
language = "English (US)",
volume = "21",
pages = "2314--2317",
journal = "PACE - Pacing and Clinical Electrophysiology",
issn = "0147-8389",
publisher = "Wiley-Blackwell",
number = "11 II",

}

TY - JOUR

T1 - Spontaneous versus extraction related injuries associated with Accufix J-wire atrial pacemaker lead

T2 - Tracking changes in patient management

AU - Kawanishi, D. T.

AU - Brinker, J. A.

AU - Reeves, R.

AU - Kay, G. N.

AU - Gross, Jay N.

AU - Pioger, G.

AU - Petitot, J. C.

AU - Esler, A.

AU - Grunkemeier, G.

PY - 1998

Y1 - 1998

N2 - To make recommendations for management of potentially fatal failure of the Accufix series of atrial J-wire permanent pacemaker leads, we closely monitored the number of injuries and fatalities resulting either from spontaneous fracture of the J-wire or from attempts to extract the lead. In a population of 30,357 patients, 2,298 patients are enrolled in a prospective follow-up Multicenter Study, the remainder are patients with known clinical status from voluntary reporting, and 2,992 patients died following implant. In the remaining 27,365 patients, 6 deaths have been attributed to J-wire related injury (J-inj) while 13 were complications (E-inj) associated with 4,076 lead extraction procedures (3,974 intravascular (intra)/102 primary thoracotomy (PT). The date of occurrences were from 1994 to November 1997. Conclusions: (1) Since lead extractions were not conducted in a controlled study, it is not known whether the deaths associated with lead extraction is in excess of what would have occurred if these leads had not been removed in this specific subset. (2) Awareness of the procedure related complication rate appears to have moderated the rate of lead extraction and may ultimately lead to management that reasonably balances the risks of patient injury.

AB - To make recommendations for management of potentially fatal failure of the Accufix series of atrial J-wire permanent pacemaker leads, we closely monitored the number of injuries and fatalities resulting either from spontaneous fracture of the J-wire or from attempts to extract the lead. In a population of 30,357 patients, 2,298 patients are enrolled in a prospective follow-up Multicenter Study, the remainder are patients with known clinical status from voluntary reporting, and 2,992 patients died following implant. In the remaining 27,365 patients, 6 deaths have been attributed to J-wire related injury (J-inj) while 13 were complications (E-inj) associated with 4,076 lead extraction procedures (3,974 intravascular (intra)/102 primary thoracotomy (PT). The date of occurrences were from 1994 to November 1997. Conclusions: (1) Since lead extractions were not conducted in a controlled study, it is not known whether the deaths associated with lead extraction is in excess of what would have occurred if these leads had not been removed in this specific subset. (2) Awareness of the procedure related complication rate appears to have moderated the rate of lead extraction and may ultimately lead to management that reasonably balances the risks of patient injury.

KW - Clinical follow-up

KW - Device failure

KW - Lead extraction

KW - Permanent pacemaker lead

UR - http://www.scopus.com/inward/record.url?scp=0031753268&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0031753268&partnerID=8YFLogxK

M3 - Article

C2 - 9825339

AN - SCOPUS:0031753268

VL - 21

SP - 2314

EP - 2317

JO - PACE - Pacing and Clinical Electrophysiology

JF - PACE - Pacing and Clinical Electrophysiology

SN - 0147-8389

IS - 11 II

ER -