Scatter radiation from chest radiographs: is there a risk to infants in a typical NICU?

Angela M. Trinh, Alan H. Schoenfeld, Terry L. Levin

Research output: Contribution to journalArticle

10 Citations (Scopus)

Abstract

OBJECTIVE: To evaluate the dose of scatter radiation to infants in a NICU in order to determine the minimal safe distance between isolettes. MATERIALS AND METHODS: Dose secondary to scattered radiation from an acrylic phantom exposed to vertical and horizontal beam exposures at 56 kVp was measured at 93 cm and 125 cm from the center of the phantom. This corresponds to 2 and 3 ft between standard isolettes, respectively. For horizontal exposures, the dosimeter was placed directly behind a CR plate and scatter dose at 90-degrees and 135-degrees from the incident beam was also measured. Exposures were obtained at 160 mAs and the results were extrapolated to correspond to 2.5 mAs. Four measurements were taken at each point and averaged. RESULTS: At 125 cm and 93 cm there was minimal scatter compared to daily natural background radiation dose (8.493 microGy). Greatest scatter dose obtained from a horizontal beam exposure at 135 degrees from the incident beam was still far below background radiation. CONCLUSION: Scatter radiation dose from a single exposure as well as cumulative scatter dose from numerous exposures is significantly below natural background radiation. Infants in neighboring isolettes are not at added risk from radiation scatter as long as the isolettes are separated by at least 2 ft.

Original languageEnglish (US)
Pages (from-to)704-707
Number of pages4
JournalPediatric Radiology
Volume40
Issue number5
StatePublished - May 2010
Externally publishedYes

Fingerprint

Background Radiation
Thorax
Radiation

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Pediatrics, Perinatology, and Child Health

Cite this

Scatter radiation from chest radiographs : is there a risk to infants in a typical NICU? / Trinh, Angela M.; Schoenfeld, Alan H.; Levin, Terry L.

In: Pediatric Radiology, Vol. 40, No. 5, 05.2010, p. 704-707.

Research output: Contribution to journalArticle

@article{764f2c10d2084e9681a5e3fb0e9a315b,
title = "Scatter radiation from chest radiographs: is there a risk to infants in a typical NICU?",
abstract = "OBJECTIVE: To evaluate the dose of scatter radiation to infants in a NICU in order to determine the minimal safe distance between isolettes. MATERIALS AND METHODS: Dose secondary to scattered radiation from an acrylic phantom exposed to vertical and horizontal beam exposures at 56 kVp was measured at 93 cm and 125 cm from the center of the phantom. This corresponds to 2 and 3 ft between standard isolettes, respectively. For horizontal exposures, the dosimeter was placed directly behind a CR plate and scatter dose at 90-degrees and 135-degrees from the incident beam was also measured. Exposures were obtained at 160 mAs and the results were extrapolated to correspond to 2.5 mAs. Four measurements were taken at each point and averaged. RESULTS: At 125 cm and 93 cm there was minimal scatter compared to daily natural background radiation dose (8.493 microGy). Greatest scatter dose obtained from a horizontal beam exposure at 135 degrees from the incident beam was still far below background radiation. CONCLUSION: Scatter radiation dose from a single exposure as well as cumulative scatter dose from numerous exposures is significantly below natural background radiation. Infants in neighboring isolettes are not at added risk from radiation scatter as long as the isolettes are separated by at least 2 ft.",
author = "Trinh, {Angela M.} and Schoenfeld, {Alan H.} and Levin, {Terry L.}",
year = "2010",
month = "5",
language = "English (US)",
volume = "40",
pages = "704--707",
journal = "Pediatric Radiology",
issn = "0301-0449",
publisher = "Springer Verlag",
number = "5",

}

TY - JOUR

T1 - Scatter radiation from chest radiographs

T2 - is there a risk to infants in a typical NICU?

AU - Trinh, Angela M.

AU - Schoenfeld, Alan H.

AU - Levin, Terry L.

PY - 2010/5

Y1 - 2010/5

N2 - OBJECTIVE: To evaluate the dose of scatter radiation to infants in a NICU in order to determine the minimal safe distance between isolettes. MATERIALS AND METHODS: Dose secondary to scattered radiation from an acrylic phantom exposed to vertical and horizontal beam exposures at 56 kVp was measured at 93 cm and 125 cm from the center of the phantom. This corresponds to 2 and 3 ft between standard isolettes, respectively. For horizontal exposures, the dosimeter was placed directly behind a CR plate and scatter dose at 90-degrees and 135-degrees from the incident beam was also measured. Exposures were obtained at 160 mAs and the results were extrapolated to correspond to 2.5 mAs. Four measurements were taken at each point and averaged. RESULTS: At 125 cm and 93 cm there was minimal scatter compared to daily natural background radiation dose (8.493 microGy). Greatest scatter dose obtained from a horizontal beam exposure at 135 degrees from the incident beam was still far below background radiation. CONCLUSION: Scatter radiation dose from a single exposure as well as cumulative scatter dose from numerous exposures is significantly below natural background radiation. Infants in neighboring isolettes are not at added risk from radiation scatter as long as the isolettes are separated by at least 2 ft.

AB - OBJECTIVE: To evaluate the dose of scatter radiation to infants in a NICU in order to determine the minimal safe distance between isolettes. MATERIALS AND METHODS: Dose secondary to scattered radiation from an acrylic phantom exposed to vertical and horizontal beam exposures at 56 kVp was measured at 93 cm and 125 cm from the center of the phantom. This corresponds to 2 and 3 ft between standard isolettes, respectively. For horizontal exposures, the dosimeter was placed directly behind a CR plate and scatter dose at 90-degrees and 135-degrees from the incident beam was also measured. Exposures were obtained at 160 mAs and the results were extrapolated to correspond to 2.5 mAs. Four measurements were taken at each point and averaged. RESULTS: At 125 cm and 93 cm there was minimal scatter compared to daily natural background radiation dose (8.493 microGy). Greatest scatter dose obtained from a horizontal beam exposure at 135 degrees from the incident beam was still far below background radiation. CONCLUSION: Scatter radiation dose from a single exposure as well as cumulative scatter dose from numerous exposures is significantly below natural background radiation. Infants in neighboring isolettes are not at added risk from radiation scatter as long as the isolettes are separated by at least 2 ft.

UR - http://www.scopus.com/inward/record.url?scp=77954529097&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=77954529097&partnerID=8YFLogxK

M3 - Article

C2 - 19997727

AN - SCOPUS:77954529097

VL - 40

SP - 704

EP - 707

JO - Pediatric Radiology

JF - Pediatric Radiology

SN - 0301-0449

IS - 5

ER -