Quality assurance of picture archiving communication systems with laser film digitizers

Peter D. Esser, Ethan J. Halpern, E. S. Amis

Research output: Contribution to journalArticle

7 Citations (Scopus)

Abstract

A comprehensive quality assurance (QA) program should be implemented for all teleradiology and picture archival and communications (PACS) systems. In this report, we summarize our QA experience with a teleradiology system that includes a laser digitizer for x-ray film. A key component required for the evaluation of laser film scanners is an appropriate test patterns; digitizers should be evaluated with enhanced test patterns specifically designed for this purpose. The phantom pattern should measure high-contrast resolution, low-contrast discrimination, gray scale linearity, geometric distortion, and noise. In addition, a uniformly exposed sheet of film (approximately 0.3 optical density) serves as a good phantom for testing screen nonuniformities of viewing station monitors. It is also suggested that clinical images should be included in a QA program. Finally, it is recommended that any discrepancies in the interpretation of teleradiology images should be monitored and investigated.

Original languageEnglish (US)
Pages (from-to)248-250
Number of pages3
JournalJournal of Digital Imaging
Volume4
Issue number4
DOIs
StatePublished - Nov 1991
Externally publishedYes

Fingerprint

Teleradiology
Radiology Information Systems
Quality assurance
Communication systems
Lasers
Motion Pictures
Density (optical)
Noise
X-Rays
X rays
Testing

ASJC Scopus subject areas

  • Radiological and Ultrasound Technology
  • Radiology Nuclear Medicine and imaging
  • Computer Science Applications
  • Computer Vision and Pattern Recognition

Cite this

Quality assurance of picture archiving communication systems with laser film digitizers. / Esser, Peter D.; Halpern, Ethan J.; Amis, E. S.

In: Journal of Digital Imaging, Vol. 4, No. 4, 11.1991, p. 248-250.

Research output: Contribution to journalArticle

Esser, Peter D. ; Halpern, Ethan J. ; Amis, E. S. / Quality assurance of picture archiving communication systems with laser film digitizers. In: Journal of Digital Imaging. 1991 ; Vol. 4, No. 4. pp. 248-250.
@article{d09fb080a45e4df2a53ce3e90f407f09,
title = "Quality assurance of picture archiving communication systems with laser film digitizers",
abstract = "A comprehensive quality assurance (QA) program should be implemented for all teleradiology and picture archival and communications (PACS) systems. In this report, we summarize our QA experience with a teleradiology system that includes a laser digitizer for x-ray film. A key component required for the evaluation of laser film scanners is an appropriate test patterns; digitizers should be evaluated with enhanced test patterns specifically designed for this purpose. The phantom pattern should measure high-contrast resolution, low-contrast discrimination, gray scale linearity, geometric distortion, and noise. In addition, a uniformly exposed sheet of film (approximately 0.3 optical density) serves as a good phantom for testing screen nonuniformities of viewing station monitors. It is also suggested that clinical images should be included in a QA program. Finally, it is recommended that any discrepancies in the interpretation of teleradiology images should be monitored and investigated.",
author = "Esser, {Peter D.} and Halpern, {Ethan J.} and Amis, {E. S.}",
year = "1991",
month = "11",
doi = "10.1007/BF03173907",
language = "English (US)",
volume = "4",
pages = "248--250",
journal = "Journal of Digital Imaging",
issn = "0897-1889",
publisher = "Springer New York",
number = "4",

}

TY - JOUR

T1 - Quality assurance of picture archiving communication systems with laser film digitizers

AU - Esser, Peter D.

AU - Halpern, Ethan J.

AU - Amis, E. S.

PY - 1991/11

Y1 - 1991/11

N2 - A comprehensive quality assurance (QA) program should be implemented for all teleradiology and picture archival and communications (PACS) systems. In this report, we summarize our QA experience with a teleradiology system that includes a laser digitizer for x-ray film. A key component required for the evaluation of laser film scanners is an appropriate test patterns; digitizers should be evaluated with enhanced test patterns specifically designed for this purpose. The phantom pattern should measure high-contrast resolution, low-contrast discrimination, gray scale linearity, geometric distortion, and noise. In addition, a uniformly exposed sheet of film (approximately 0.3 optical density) serves as a good phantom for testing screen nonuniformities of viewing station monitors. It is also suggested that clinical images should be included in a QA program. Finally, it is recommended that any discrepancies in the interpretation of teleradiology images should be monitored and investigated.

AB - A comprehensive quality assurance (QA) program should be implemented for all teleradiology and picture archival and communications (PACS) systems. In this report, we summarize our QA experience with a teleradiology system that includes a laser digitizer for x-ray film. A key component required for the evaluation of laser film scanners is an appropriate test patterns; digitizers should be evaluated with enhanced test patterns specifically designed for this purpose. The phantom pattern should measure high-contrast resolution, low-contrast discrimination, gray scale linearity, geometric distortion, and noise. In addition, a uniformly exposed sheet of film (approximately 0.3 optical density) serves as a good phantom for testing screen nonuniformities of viewing station monitors. It is also suggested that clinical images should be included in a QA program. Finally, it is recommended that any discrepancies in the interpretation of teleradiology images should be monitored and investigated.

UR - http://www.scopus.com/inward/record.url?scp=0026248119&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0026248119&partnerID=8YFLogxK

U2 - 10.1007/BF03173907

DO - 10.1007/BF03173907

M3 - Article

VL - 4

SP - 248

EP - 250

JO - Journal of Digital Imaging

JF - Journal of Digital Imaging

SN - 0897-1889

IS - 4

ER -