Quality assurance of an image guided intracranial stereotactic positioning system for radiosurgery treatment with helical tomotherapy

Emilie T. Soisson, Nick Hardcastle, Wolfgang A. Tome

Research output: Contribution to journalArticle

8 Citations (Scopus)

Abstract

The aim of this work was to determine the accuracy and precision of stereotactic localization and treatment delivery using a helical tomotherapy based stereotactic radiosurgery (SRS) system. A tomotherapy specific radiosurgery workflow was designed that exploits the system's on board megavotage CT (MVCT) imaging system so that it not only provides a pre-treatment volumetric verification image that can be used for stereotactic localization, eliminating the need for a patient-frame based coordinate system, but also supplies the treatment planning image. Using an imaging guidance based intracranial stereotactic positioning system, a head ring and tabletop docking device are used only for fixation, while image guidance is used for localization. Due to the unconventional workflow, a methodology for determining the localization accuracy was developed and results were compared to other linear accelerator based radiosurgery systems. In this work, the localization error using volumetric localization was found to be 0.45 mm ± 0.17 mm, indicating a localization precision of 0.3 mm within a 95% confidence interval. In addition, procedures for testing the delivery accuracy of the Tomotherapy system are described. Results show that the accuracy of the delivery can be verified to within ±1 voxel dimension. These results are well within conventional SRS tolerances and compare favorably to other linear accelerator based techniques.

Original languageEnglish (US)
Pages (from-to)277-285
Number of pages9
JournalJournal of Neuro-Oncology
Volume98
Issue number2
DOIs
StatePublished - Jun 2010
Externally publishedYes

Fingerprint

Intensity-Modulated Radiotherapy
Radiosurgery
Workflow
Particle Accelerators
Therapeutics
Head
Confidence Intervals
Equipment and Supplies

Keywords

  • Image guidance
  • Stereotactic localization
  • Stereotactic radiosurgery
  • Tomotherapy

ASJC Scopus subject areas

  • Clinical Neurology
  • Cancer Research
  • Oncology
  • Neurology

Cite this

Quality assurance of an image guided intracranial stereotactic positioning system for radiosurgery treatment with helical tomotherapy. / Soisson, Emilie T.; Hardcastle, Nick; Tome, Wolfgang A.

In: Journal of Neuro-Oncology, Vol. 98, No. 2, 06.2010, p. 277-285.

Research output: Contribution to journalArticle

@article{182c28004aec40bc8c1b83ee64e6aa55,
title = "Quality assurance of an image guided intracranial stereotactic positioning system for radiosurgery treatment with helical tomotherapy",
abstract = "The aim of this work was to determine the accuracy and precision of stereotactic localization and treatment delivery using a helical tomotherapy based stereotactic radiosurgery (SRS) system. A tomotherapy specific radiosurgery workflow was designed that exploits the system's on board megavotage CT (MVCT) imaging system so that it not only provides a pre-treatment volumetric verification image that can be used for stereotactic localization, eliminating the need for a patient-frame based coordinate system, but also supplies the treatment planning image. Using an imaging guidance based intracranial stereotactic positioning system, a head ring and tabletop docking device are used only for fixation, while image guidance is used for localization. Due to the unconventional workflow, a methodology for determining the localization accuracy was developed and results were compared to other linear accelerator based radiosurgery systems. In this work, the localization error using volumetric localization was found to be 0.45 mm ± 0.17 mm, indicating a localization precision of 0.3 mm within a 95{\%} confidence interval. In addition, procedures for testing the delivery accuracy of the Tomotherapy system are described. Results show that the accuracy of the delivery can be verified to within ±1 voxel dimension. These results are well within conventional SRS tolerances and compare favorably to other linear accelerator based techniques.",
keywords = "Image guidance, Stereotactic localization, Stereotactic radiosurgery, Tomotherapy",
author = "Soisson, {Emilie T.} and Nick Hardcastle and Tome, {Wolfgang A.}",
year = "2010",
month = "6",
doi = "10.1007/s11060-010-0227-4",
language = "English (US)",
volume = "98",
pages = "277--285",
journal = "Journal of Neuro-Oncology",
issn = "0167-594X",
publisher = "Kluwer Academic Publishers",
number = "2",

}

TY - JOUR

T1 - Quality assurance of an image guided intracranial stereotactic positioning system for radiosurgery treatment with helical tomotherapy

AU - Soisson, Emilie T.

AU - Hardcastle, Nick

AU - Tome, Wolfgang A.

PY - 2010/6

Y1 - 2010/6

N2 - The aim of this work was to determine the accuracy and precision of stereotactic localization and treatment delivery using a helical tomotherapy based stereotactic radiosurgery (SRS) system. A tomotherapy specific radiosurgery workflow was designed that exploits the system's on board megavotage CT (MVCT) imaging system so that it not only provides a pre-treatment volumetric verification image that can be used for stereotactic localization, eliminating the need for a patient-frame based coordinate system, but also supplies the treatment planning image. Using an imaging guidance based intracranial stereotactic positioning system, a head ring and tabletop docking device are used only for fixation, while image guidance is used for localization. Due to the unconventional workflow, a methodology for determining the localization accuracy was developed and results were compared to other linear accelerator based radiosurgery systems. In this work, the localization error using volumetric localization was found to be 0.45 mm ± 0.17 mm, indicating a localization precision of 0.3 mm within a 95% confidence interval. In addition, procedures for testing the delivery accuracy of the Tomotherapy system are described. Results show that the accuracy of the delivery can be verified to within ±1 voxel dimension. These results are well within conventional SRS tolerances and compare favorably to other linear accelerator based techniques.

AB - The aim of this work was to determine the accuracy and precision of stereotactic localization and treatment delivery using a helical tomotherapy based stereotactic radiosurgery (SRS) system. A tomotherapy specific radiosurgery workflow was designed that exploits the system's on board megavotage CT (MVCT) imaging system so that it not only provides a pre-treatment volumetric verification image that can be used for stereotactic localization, eliminating the need for a patient-frame based coordinate system, but also supplies the treatment planning image. Using an imaging guidance based intracranial stereotactic positioning system, a head ring and tabletop docking device are used only for fixation, while image guidance is used for localization. Due to the unconventional workflow, a methodology for determining the localization accuracy was developed and results were compared to other linear accelerator based radiosurgery systems. In this work, the localization error using volumetric localization was found to be 0.45 mm ± 0.17 mm, indicating a localization precision of 0.3 mm within a 95% confidence interval. In addition, procedures for testing the delivery accuracy of the Tomotherapy system are described. Results show that the accuracy of the delivery can be verified to within ±1 voxel dimension. These results are well within conventional SRS tolerances and compare favorably to other linear accelerator based techniques.

KW - Image guidance

KW - Stereotactic localization

KW - Stereotactic radiosurgery

KW - Tomotherapy

UR - http://www.scopus.com/inward/record.url?scp=77953618801&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=77953618801&partnerID=8YFLogxK

U2 - 10.1007/s11060-010-0227-4

DO - 10.1007/s11060-010-0227-4

M3 - Article

C2 - 20505980

AN - SCOPUS:77953618801

VL - 98

SP - 277

EP - 285

JO - Journal of Neuro-Oncology

JF - Journal of Neuro-Oncology

SN - 0167-594X

IS - 2

ER -