Quality assurance of an image guided intracranial stereotactic positioning system

Emilie T. Soisson, Geoff Sobering, Daniel Lucas, Edward Chao, Gustavo Olivera, Wolfgang A. Tome

Research output: Contribution to journalArticle

5 Citations (Scopus)

Abstract

This work reports on the development and testing of an intracranial stereotactic patient positioning system (ISPPS) for Tomotherapy. The ISPPS consists of the combination of a head frame, head frame couch interface (HCI), megavoltage CT (MVCT), and optical tracking camera system. Three quality assurance tests were designed to quantify the positioning system's ability to localize an intracranial target. The first two of these tests were designed to determine (a) the ability of the MVCT to detect a known shift applied to an anthropomorphic phantom and (b) the precision of fixing the phantom to the treatment couch via a head frame and specially designed head frame couch interface. A system verification test, using a phantom and EDR2 film, was used to determine the overall delivery precision through comparison of a measured dose distribution on film to calculated dose. The average net translational difference between a known shift applied to a phantom and that detected by MVCT image fusion was 0.62 mm. Setup reproducibility of the head frame was measured with both MVCT and optical tracking. The frame setup precision was found to be well within 1 mm for translations as well as rotations. A system delivery verification test in phantom using film showed spatial agreement between planned and delivered dose distributions to within 1 mm.

Original languageEnglish (US)
Pages (from-to)39-49
Number of pages11
JournalTechnology in Cancer Research and Treatment
Volume8
Issue number1
StatePublished - Feb 2009
Externally publishedYes

Fingerprint

Head
Patient Positioning
Therapeutics

Keywords

  • Intracranial localization
  • MVCT
  • Stereotactic radiosurgery
  • Tomotherapy

ASJC Scopus subject areas

  • Cancer Research
  • Oncology

Cite this

Quality assurance of an image guided intracranial stereotactic positioning system. / Soisson, Emilie T.; Sobering, Geoff; Lucas, Daniel; Chao, Edward; Olivera, Gustavo; Tome, Wolfgang A.

In: Technology in Cancer Research and Treatment, Vol. 8, No. 1, 02.2009, p. 39-49.

Research output: Contribution to journalArticle

Soisson, ET, Sobering, G, Lucas, D, Chao, E, Olivera, G & Tome, WA 2009, 'Quality assurance of an image guided intracranial stereotactic positioning system', Technology in Cancer Research and Treatment, vol. 8, no. 1, pp. 39-49.
Soisson, Emilie T. ; Sobering, Geoff ; Lucas, Daniel ; Chao, Edward ; Olivera, Gustavo ; Tome, Wolfgang A. / Quality assurance of an image guided intracranial stereotactic positioning system. In: Technology in Cancer Research and Treatment. 2009 ; Vol. 8, No. 1. pp. 39-49.
@article{53d7b117f016454cb70c57b51e283377,
title = "Quality assurance of an image guided intracranial stereotactic positioning system",
abstract = "This work reports on the development and testing of an intracranial stereotactic patient positioning system (ISPPS) for Tomotherapy. The ISPPS consists of the combination of a head frame, head frame couch interface (HCI), megavoltage CT (MVCT), and optical tracking camera system. Three quality assurance tests were designed to quantify the positioning system's ability to localize an intracranial target. The first two of these tests were designed to determine (a) the ability of the MVCT to detect a known shift applied to an anthropomorphic phantom and (b) the precision of fixing the phantom to the treatment couch via a head frame and specially designed head frame couch interface. A system verification test, using a phantom and EDR2 film, was used to determine the overall delivery precision through comparison of a measured dose distribution on film to calculated dose. The average net translational difference between a known shift applied to a phantom and that detected by MVCT image fusion was 0.62 mm. Setup reproducibility of the head frame was measured with both MVCT and optical tracking. The frame setup precision was found to be well within 1 mm for translations as well as rotations. A system delivery verification test in phantom using film showed spatial agreement between planned and delivered dose distributions to within 1 mm.",
keywords = "Intracranial localization, MVCT, Stereotactic radiosurgery, Tomotherapy",
author = "Soisson, {Emilie T.} and Geoff Sobering and Daniel Lucas and Edward Chao and Gustavo Olivera and Tome, {Wolfgang A.}",
year = "2009",
month = "2",
language = "English (US)",
volume = "8",
pages = "39--49",
journal = "Technology in Cancer Research and Treatment",
issn = "1533-0346",
publisher = "Adenine Press",
number = "1",

}

TY - JOUR

T1 - Quality assurance of an image guided intracranial stereotactic positioning system

AU - Soisson, Emilie T.

AU - Sobering, Geoff

AU - Lucas, Daniel

AU - Chao, Edward

AU - Olivera, Gustavo

AU - Tome, Wolfgang A.

PY - 2009/2

Y1 - 2009/2

N2 - This work reports on the development and testing of an intracranial stereotactic patient positioning system (ISPPS) for Tomotherapy. The ISPPS consists of the combination of a head frame, head frame couch interface (HCI), megavoltage CT (MVCT), and optical tracking camera system. Three quality assurance tests were designed to quantify the positioning system's ability to localize an intracranial target. The first two of these tests were designed to determine (a) the ability of the MVCT to detect a known shift applied to an anthropomorphic phantom and (b) the precision of fixing the phantom to the treatment couch via a head frame and specially designed head frame couch interface. A system verification test, using a phantom and EDR2 film, was used to determine the overall delivery precision through comparison of a measured dose distribution on film to calculated dose. The average net translational difference between a known shift applied to a phantom and that detected by MVCT image fusion was 0.62 mm. Setup reproducibility of the head frame was measured with both MVCT and optical tracking. The frame setup precision was found to be well within 1 mm for translations as well as rotations. A system delivery verification test in phantom using film showed spatial agreement between planned and delivered dose distributions to within 1 mm.

AB - This work reports on the development and testing of an intracranial stereotactic patient positioning system (ISPPS) for Tomotherapy. The ISPPS consists of the combination of a head frame, head frame couch interface (HCI), megavoltage CT (MVCT), and optical tracking camera system. Three quality assurance tests were designed to quantify the positioning system's ability to localize an intracranial target. The first two of these tests were designed to determine (a) the ability of the MVCT to detect a known shift applied to an anthropomorphic phantom and (b) the precision of fixing the phantom to the treatment couch via a head frame and specially designed head frame couch interface. A system verification test, using a phantom and EDR2 film, was used to determine the overall delivery precision through comparison of a measured dose distribution on film to calculated dose. The average net translational difference between a known shift applied to a phantom and that detected by MVCT image fusion was 0.62 mm. Setup reproducibility of the head frame was measured with both MVCT and optical tracking. The frame setup precision was found to be well within 1 mm for translations as well as rotations. A system delivery verification test in phantom using film showed spatial agreement between planned and delivered dose distributions to within 1 mm.

KW - Intracranial localization

KW - MVCT

KW - Stereotactic radiosurgery

KW - Tomotherapy

UR - http://www.scopus.com/inward/record.url?scp=60849121419&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=60849121419&partnerID=8YFLogxK

M3 - Article

VL - 8

SP - 39

EP - 49

JO - Technology in Cancer Research and Treatment

JF - Technology in Cancer Research and Treatment

SN - 1533-0346

IS - 1

ER -