O-arm failure traps patient prone on a Jackson table

Sergey V. Pisklakov, John K. Houten, Murray Echt, Angelika Kosse, Apolonia E. Abramowicz

Research output: Contribution to journalArticle

Abstract

The O-arm Surgical Imaging System (Medtronic Sofamor Danek, Memphis, TN, USA) is an increasingly popular tool for spinal surgical procedures that allows for intraoperative acquisition of multi-dimensional spinal imaging that can be used either to confirm placement of spinal instrumentation or to guide spinal screw placement in conjunction with a navigation system. The machine is typically placed open from the side of the patient positioned prone on the Jackson spinal table (Mizuho OSI, Union City, CA, USA) and then closed to complete a ring around the patient to acquire images. A failure of the O-arm opening mechanism can lead to a patient becoming effectively trapped within the device with limited physician access to the patient's body, a situation that may be remedied by using a manual opening procedure. This report highlights a patient safety issue encountered when a mechanical failure of the O-arm occurred and one of the two tools required for its manual opening was missing. We describe the improvised method employed to successfully open the device.

Original languageEnglish (US)
Pages (from-to)1365-1366
Number of pages2
JournalJournal of Clinical Neuroscience
Volume22
Issue number8
DOIs
StatePublished - Aug 1 2015

Fingerprint

Equipment and Supplies
Patient Safety
Physicians

Keywords

  • Image-guided surgery
  • Navigation
  • O-arm
  • Patient safety
  • Pedicle screw
  • Spinal surgery

ASJC Scopus subject areas

  • Clinical Neurology
  • Neurology
  • Physiology (medical)

Cite this

O-arm failure traps patient prone on a Jackson table. / Pisklakov, Sergey V.; Houten, John K.; Echt, Murray; Kosse, Angelika; Abramowicz, Apolonia E.

In: Journal of Clinical Neuroscience, Vol. 22, No. 8, 01.08.2015, p. 1365-1366.

Research output: Contribution to journalArticle

Pisklakov, Sergey V. ; Houten, John K. ; Echt, Murray ; Kosse, Angelika ; Abramowicz, Apolonia E. / O-arm failure traps patient prone on a Jackson table. In: Journal of Clinical Neuroscience. 2015 ; Vol. 22, No. 8. pp. 1365-1366.
@article{3f9cbcb4cc9d45f8b132cc2a2fc79997,
title = "O-arm failure traps patient prone on a Jackson table",
abstract = "The O-arm Surgical Imaging System (Medtronic Sofamor Danek, Memphis, TN, USA) is an increasingly popular tool for spinal surgical procedures that allows for intraoperative acquisition of multi-dimensional spinal imaging that can be used either to confirm placement of spinal instrumentation or to guide spinal screw placement in conjunction with a navigation system. The machine is typically placed open from the side of the patient positioned prone on the Jackson spinal table (Mizuho OSI, Union City, CA, USA) and then closed to complete a ring around the patient to acquire images. A failure of the O-arm opening mechanism can lead to a patient becoming effectively trapped within the device with limited physician access to the patient's body, a situation that may be remedied by using a manual opening procedure. This report highlights a patient safety issue encountered when a mechanical failure of the O-arm occurred and one of the two tools required for its manual opening was missing. We describe the improvised method employed to successfully open the device.",
keywords = "Image-guided surgery, Navigation, O-arm, Patient safety, Pedicle screw, Spinal surgery",
author = "Pisklakov, {Sergey V.} and Houten, {John K.} and Murray Echt and Angelika Kosse and Abramowicz, {Apolonia E.}",
year = "2015",
month = "8",
day = "1",
doi = "10.1016/j.jocn.2015.03.003",
language = "English (US)",
volume = "22",
pages = "1365--1366",
journal = "Journal of Clinical Neuroscience",
issn = "0967-5868",
publisher = "Churchill Livingstone",
number = "8",

}

TY - JOUR

T1 - O-arm failure traps patient prone on a Jackson table

AU - Pisklakov, Sergey V.

AU - Houten, John K.

AU - Echt, Murray

AU - Kosse, Angelika

AU - Abramowicz, Apolonia E.

PY - 2015/8/1

Y1 - 2015/8/1

N2 - The O-arm Surgical Imaging System (Medtronic Sofamor Danek, Memphis, TN, USA) is an increasingly popular tool for spinal surgical procedures that allows for intraoperative acquisition of multi-dimensional spinal imaging that can be used either to confirm placement of spinal instrumentation or to guide spinal screw placement in conjunction with a navigation system. The machine is typically placed open from the side of the patient positioned prone on the Jackson spinal table (Mizuho OSI, Union City, CA, USA) and then closed to complete a ring around the patient to acquire images. A failure of the O-arm opening mechanism can lead to a patient becoming effectively trapped within the device with limited physician access to the patient's body, a situation that may be remedied by using a manual opening procedure. This report highlights a patient safety issue encountered when a mechanical failure of the O-arm occurred and one of the two tools required for its manual opening was missing. We describe the improvised method employed to successfully open the device.

AB - The O-arm Surgical Imaging System (Medtronic Sofamor Danek, Memphis, TN, USA) is an increasingly popular tool for spinal surgical procedures that allows for intraoperative acquisition of multi-dimensional spinal imaging that can be used either to confirm placement of spinal instrumentation or to guide spinal screw placement in conjunction with a navigation system. The machine is typically placed open from the side of the patient positioned prone on the Jackson spinal table (Mizuho OSI, Union City, CA, USA) and then closed to complete a ring around the patient to acquire images. A failure of the O-arm opening mechanism can lead to a patient becoming effectively trapped within the device with limited physician access to the patient's body, a situation that may be remedied by using a manual opening procedure. This report highlights a patient safety issue encountered when a mechanical failure of the O-arm occurred and one of the two tools required for its manual opening was missing. We describe the improvised method employed to successfully open the device.

KW - Image-guided surgery

KW - Navigation

KW - O-arm

KW - Patient safety

KW - Pedicle screw

KW - Spinal surgery

UR - http://www.scopus.com/inward/record.url?scp=84937513985&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84937513985&partnerID=8YFLogxK

U2 - 10.1016/j.jocn.2015.03.003

DO - 10.1016/j.jocn.2015.03.003

M3 - Article

C2 - 25937572

AN - SCOPUS:84937513985

VL - 22

SP - 1365

EP - 1366

JO - Journal of Clinical Neuroscience

JF - Journal of Clinical Neuroscience

SN - 0967-5868

IS - 8

ER -