Abstract

Purpose: To introduce a new power injection technology that generates data as a digital imaging and communications in medicine (DICOM) image linked to individual patient imaging studies. In addition, to determine the fraction of patients in a subject cohort whose contrast injection data was captured as a DICOM image and to analyze contrast injection properties for those patients. Methods: Over a 1-month period, authors performed a retrospective evaluation of 242 patients' consecutive contrastenhanced computed tomography (CT) studies from a single 320-detector CT scanner in an academic radiology department. Authors gathered unique patient and examination identifiers, prescribed and injected contrast and saline volumes, prescribed and injected flow rate, and mean and maximum injection pressures. The literature was reviewed for the initial description of power injectors in radiology. Results: Of the 242 CT studies evaluated, 98% had contrast injection data amended to the radiology images. For all patients, the mean volume of residual contrast was 5 mL. The differences between the prescribed and actual flow rate were small. Three patients reached the maximum pressure of 300 psi. There were no contrast extravasations. Discussion: The most clinically relevant finding was that the injector and software system generated a detailed report of contrast administration. In 98% of the cases, this report was incorporated into the patient's permanent medical record and was available to the radiologist via a single DICOM image. Conclusion: Contrast injection data can be captured in DICOM format and reliably attached to a clinical contrastenhanced CT image set for radiologist use.

Original languageEnglish (US)
Pages (from-to)120-125
Number of pages6
JournalRadiologic Technology
Volume84
Issue number2
StatePublished - Nov 1 2012
Externally publishedYes

Fingerprint

Technology
Injections
Communication
Radiology
Medicine
Tomography
X-Ray Computed Tomography Scanners
Pressure
Residual Volume
Medical Records
Software
Radiologists
Power (Psychology)

ASJC Scopus subject areas

  • Radiological and Ultrasound Technology
  • Radiology Nuclear Medicine and imaging

Cite this

Rybicki, F. J., Piazzo, K., Prior, R., Wake, N., & Dill, K. E. (2012). Iodinated contrast injection data from a new technology. Radiologic Technology, 84(2), 120-125.

Iodinated contrast injection data from a new technology. / Rybicki, Frank J.; Piazzo, Kathleen; Prior, Richard; Wake, Nicole; Dill, Karin E.

In: Radiologic Technology, Vol. 84, No. 2, 01.11.2012, p. 120-125.

Research output: Contribution to journalReview article

Rybicki, FJ, Piazzo, K, Prior, R, Wake, N & Dill, KE 2012, 'Iodinated contrast injection data from a new technology', Radiologic Technology, vol. 84, no. 2, pp. 120-125.
Rybicki FJ, Piazzo K, Prior R, Wake N, Dill KE. Iodinated contrast injection data from a new technology. Radiologic Technology. 2012 Nov 1;84(2):120-125.
Rybicki, Frank J. ; Piazzo, Kathleen ; Prior, Richard ; Wake, Nicole ; Dill, Karin E. / Iodinated contrast injection data from a new technology. In: Radiologic Technology. 2012 ; Vol. 84, No. 2. pp. 120-125.
@article{bb7e86e5c8934fd49f92a0890267cc5c,
title = "Iodinated contrast injection data from a new technology",
abstract = "Purpose: To introduce a new power injection technology that generates data as a digital imaging and communications in medicine (DICOM) image linked to individual patient imaging studies. In addition, to determine the fraction of patients in a subject cohort whose contrast injection data was captured as a DICOM image and to analyze contrast injection properties for those patients. Methods: Over a 1-month period, authors performed a retrospective evaluation of 242 patients' consecutive contrastenhanced computed tomography (CT) studies from a single 320-detector CT scanner in an academic radiology department. Authors gathered unique patient and examination identifiers, prescribed and injected contrast and saline volumes, prescribed and injected flow rate, and mean and maximum injection pressures. The literature was reviewed for the initial description of power injectors in radiology. Results: Of the 242 CT studies evaluated, 98{\%} had contrast injection data amended to the radiology images. For all patients, the mean volume of residual contrast was 5 mL. The differences between the prescribed and actual flow rate were small. Three patients reached the maximum pressure of 300 psi. There were no contrast extravasations. Discussion: The most clinically relevant finding was that the injector and software system generated a detailed report of contrast administration. In 98{\%} of the cases, this report was incorporated into the patient's permanent medical record and was available to the radiologist via a single DICOM image. Conclusion: Contrast injection data can be captured in DICOM format and reliably attached to a clinical contrastenhanced CT image set for radiologist use.",
author = "Rybicki, {Frank J.} and Kathleen Piazzo and Richard Prior and Nicole Wake and Dill, {Karin E.}",
year = "2012",
month = "11",
day = "1",
language = "English (US)",
volume = "84",
pages = "120--125",
journal = "Radiologic Technology",
issn = "0033-8397",
publisher = "Grune And Stratton",
number = "2",

}

TY - JOUR

T1 - Iodinated contrast injection data from a new technology

AU - Rybicki, Frank J.

AU - Piazzo, Kathleen

AU - Prior, Richard

AU - Wake, Nicole

AU - Dill, Karin E.

PY - 2012/11/1

Y1 - 2012/11/1

N2 - Purpose: To introduce a new power injection technology that generates data as a digital imaging and communications in medicine (DICOM) image linked to individual patient imaging studies. In addition, to determine the fraction of patients in a subject cohort whose contrast injection data was captured as a DICOM image and to analyze contrast injection properties for those patients. Methods: Over a 1-month period, authors performed a retrospective evaluation of 242 patients' consecutive contrastenhanced computed tomography (CT) studies from a single 320-detector CT scanner in an academic radiology department. Authors gathered unique patient and examination identifiers, prescribed and injected contrast and saline volumes, prescribed and injected flow rate, and mean and maximum injection pressures. The literature was reviewed for the initial description of power injectors in radiology. Results: Of the 242 CT studies evaluated, 98% had contrast injection data amended to the radiology images. For all patients, the mean volume of residual contrast was 5 mL. The differences between the prescribed and actual flow rate were small. Three patients reached the maximum pressure of 300 psi. There were no contrast extravasations. Discussion: The most clinically relevant finding was that the injector and software system generated a detailed report of contrast administration. In 98% of the cases, this report was incorporated into the patient's permanent medical record and was available to the radiologist via a single DICOM image. Conclusion: Contrast injection data can be captured in DICOM format and reliably attached to a clinical contrastenhanced CT image set for radiologist use.

AB - Purpose: To introduce a new power injection technology that generates data as a digital imaging and communications in medicine (DICOM) image linked to individual patient imaging studies. In addition, to determine the fraction of patients in a subject cohort whose contrast injection data was captured as a DICOM image and to analyze contrast injection properties for those patients. Methods: Over a 1-month period, authors performed a retrospective evaluation of 242 patients' consecutive contrastenhanced computed tomography (CT) studies from a single 320-detector CT scanner in an academic radiology department. Authors gathered unique patient and examination identifiers, prescribed and injected contrast and saline volumes, prescribed and injected flow rate, and mean and maximum injection pressures. The literature was reviewed for the initial description of power injectors in radiology. Results: Of the 242 CT studies evaluated, 98% had contrast injection data amended to the radiology images. For all patients, the mean volume of residual contrast was 5 mL. The differences between the prescribed and actual flow rate were small. Three patients reached the maximum pressure of 300 psi. There were no contrast extravasations. Discussion: The most clinically relevant finding was that the injector and software system generated a detailed report of contrast administration. In 98% of the cases, this report was incorporated into the patient's permanent medical record and was available to the radiologist via a single DICOM image. Conclusion: Contrast injection data can be captured in DICOM format and reliably attached to a clinical contrastenhanced CT image set for radiologist use.

UR - http://www.scopus.com/inward/record.url?scp=84877091062&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84877091062&partnerID=8YFLogxK

M3 - Review article

C2 - 23125373

AN - SCOPUS:84877091062

VL - 84

SP - 120

EP - 125

JO - Radiologic Technology

JF - Radiologic Technology

SN - 0033-8397

IS - 2

ER -