Implantable cardioverter-defibrillator. Evaluation of clinical neurologic outcome and electroencephalographic changes during implantation

David C. Adams, Eric J. Heyer, Ronald G. Emerson, Henry M. Spotnitz, Ellise S. Delphin, Christine Turner, Mitchell F. Berman

Research output: Contribution to journalArticle

30 Citations (Scopus)

Abstract

During placement of implantable cardioverter-defibrillators, ventricular arrhythmias are induced to test the function of the devices. Although cerebral hypoperfusion and ischemic electroencephalographic changes occur in patients while implantable cardioverter-defibrillators are being tested, no investigation has assessed neurologic outcome in these patients. Nine patients having either implantation or change of an implantable cardioverter-defibrillator underwent neurologic examination and neuropsychometric tests before and after the operation. After induction of general anesthesia and insertion of implantable cardioverter-defibrillator leads (when needed), ventricular fibrillation, ventricular flutter, or ventricular tachycardia, was induced by means of programmed electrical stimulation. Implantable cardioverter-defibrillator testing continued until satisfactory lead placement was confirmed. The intraoperative electroencephalographic recording was analyzed for evidence of ischemic change. In all, an electroencephalogram was recorded during 50 periods of circulatory arrest. Mean duration of the arrest periods was 13.6 seconds. By means of conventional visual inspection of the raw electroencephalogram, high-amplitude rhythmic delta or theta, voltage attenuation, or loss of fast frequency activity was observed in 30 of the arrests. By means of an automated technique of electroencephalographic interpretation based on power spectral analysis, electroencephalographic changes were correctly identified in 26 of the arrests. The incidence of these electroencephalographic changes was dependent on the arrest duration. The mean interval from arrest onset to electroencephalographic change was 7.5 seconds (standard deviation ± 1.8 seconds). In patients with electroencephalographic changes during multiple arrests, no downward trend in this interval was detected in later arrests and no evidence of persistent ischemic change was observed in electroencephalograms recorded after the conclusion of implantable cardioverter-defibrillator testing. Postoperative neurologic and neuropsychometric testing was completed in eight patients, none of whom exhibited a new neurologic deficit, exacerbation of a preexisting neurologic condition, or significant deterioration in neuropsychometric performance. We conclude that the brief arrest of cerebral circulation induced during insertion of an implantable cardioverter-defibrillator is not associated with permanent neurologic injury. (J THORAC CARDIOVASC SURG 1995; 109: 565-73).

Original languageEnglish (US)
Pages (from-to)565-573
Number of pages9
JournalThe Journal of thoracic and cardiovascular surgery
Volume109
Issue number3
DOIs
StatePublished - 1995
Externally publishedYes

Fingerprint

Implantable Defibrillators
Nervous System
Electroencephalography
Cerebrovascular Circulation
Ventricular Flutter
Nervous System Trauma
Preexisting Condition Coverage
Neurologic Examination
Ventricular Fibrillation
Ventricular Tachycardia
Neurologic Manifestations
General Anesthesia
Electric Stimulation
Cardiac Arrhythmias
Equipment and Supplies
Incidence

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine
  • Pulmonary and Respiratory Medicine
  • Surgery

Cite this

Implantable cardioverter-defibrillator. Evaluation of clinical neurologic outcome and electroencephalographic changes during implantation. / Adams, David C.; Heyer, Eric J.; Emerson, Ronald G.; Spotnitz, Henry M.; Delphin, Ellise S.; Turner, Christine; Berman, Mitchell F.

In: The Journal of thoracic and cardiovascular surgery, Vol. 109, No. 3, 1995, p. 565-573.

Research output: Contribution to journalArticle

@article{05b61514f33848efa3a3f84705dd02c3,
title = "Implantable cardioverter-defibrillator. Evaluation of clinical neurologic outcome and electroencephalographic changes during implantation",
abstract = "During placement of implantable cardioverter-defibrillators, ventricular arrhythmias are induced to test the function of the devices. Although cerebral hypoperfusion and ischemic electroencephalographic changes occur in patients while implantable cardioverter-defibrillators are being tested, no investigation has assessed neurologic outcome in these patients. Nine patients having either implantation or change of an implantable cardioverter-defibrillator underwent neurologic examination and neuropsychometric tests before and after the operation. After induction of general anesthesia and insertion of implantable cardioverter-defibrillator leads (when needed), ventricular fibrillation, ventricular flutter, or ventricular tachycardia, was induced by means of programmed electrical stimulation. Implantable cardioverter-defibrillator testing continued until satisfactory lead placement was confirmed. The intraoperative electroencephalographic recording was analyzed for evidence of ischemic change. In all, an electroencephalogram was recorded during 50 periods of circulatory arrest. Mean duration of the arrest periods was 13.6 seconds. By means of conventional visual inspection of the raw electroencephalogram, high-amplitude rhythmic delta or theta, voltage attenuation, or loss of fast frequency activity was observed in 30 of the arrests. By means of an automated technique of electroencephalographic interpretation based on power spectral analysis, electroencephalographic changes were correctly identified in 26 of the arrests. The incidence of these electroencephalographic changes was dependent on the arrest duration. The mean interval from arrest onset to electroencephalographic change was 7.5 seconds (standard deviation ± 1.8 seconds). In patients with electroencephalographic changes during multiple arrests, no downward trend in this interval was detected in later arrests and no evidence of persistent ischemic change was observed in electroencephalograms recorded after the conclusion of implantable cardioverter-defibrillator testing. Postoperative neurologic and neuropsychometric testing was completed in eight patients, none of whom exhibited a new neurologic deficit, exacerbation of a preexisting neurologic condition, or significant deterioration in neuropsychometric performance. We conclude that the brief arrest of cerebral circulation induced during insertion of an implantable cardioverter-defibrillator is not associated with permanent neurologic injury. (J THORAC CARDIOVASC SURG 1995; 109: 565-73).",
author = "Adams, {David C.} and Heyer, {Eric J.} and Emerson, {Ronald G.} and Spotnitz, {Henry M.} and Delphin, {Ellise S.} and Christine Turner and Berman, {Mitchell F.}",
year = "1995",
doi = "10.1016/S0022-5223(95)70290-3",
language = "English (US)",
volume = "109",
pages = "565--573",
journal = "Journal of Thoracic and Cardiovascular Surgery",
issn = "0022-5223",
publisher = "Mosby Inc.",
number = "3",

}

TY - JOUR

T1 - Implantable cardioverter-defibrillator. Evaluation of clinical neurologic outcome and electroencephalographic changes during implantation

AU - Adams, David C.

AU - Heyer, Eric J.

AU - Emerson, Ronald G.

AU - Spotnitz, Henry M.

AU - Delphin, Ellise S.

AU - Turner, Christine

AU - Berman, Mitchell F.

PY - 1995

Y1 - 1995

N2 - During placement of implantable cardioverter-defibrillators, ventricular arrhythmias are induced to test the function of the devices. Although cerebral hypoperfusion and ischemic electroencephalographic changes occur in patients while implantable cardioverter-defibrillators are being tested, no investigation has assessed neurologic outcome in these patients. Nine patients having either implantation or change of an implantable cardioverter-defibrillator underwent neurologic examination and neuropsychometric tests before and after the operation. After induction of general anesthesia and insertion of implantable cardioverter-defibrillator leads (when needed), ventricular fibrillation, ventricular flutter, or ventricular tachycardia, was induced by means of programmed electrical stimulation. Implantable cardioverter-defibrillator testing continued until satisfactory lead placement was confirmed. The intraoperative electroencephalographic recording was analyzed for evidence of ischemic change. In all, an electroencephalogram was recorded during 50 periods of circulatory arrest. Mean duration of the arrest periods was 13.6 seconds. By means of conventional visual inspection of the raw electroencephalogram, high-amplitude rhythmic delta or theta, voltage attenuation, or loss of fast frequency activity was observed in 30 of the arrests. By means of an automated technique of electroencephalographic interpretation based on power spectral analysis, electroencephalographic changes were correctly identified in 26 of the arrests. The incidence of these electroencephalographic changes was dependent on the arrest duration. The mean interval from arrest onset to electroencephalographic change was 7.5 seconds (standard deviation ± 1.8 seconds). In patients with electroencephalographic changes during multiple arrests, no downward trend in this interval was detected in later arrests and no evidence of persistent ischemic change was observed in electroencephalograms recorded after the conclusion of implantable cardioverter-defibrillator testing. Postoperative neurologic and neuropsychometric testing was completed in eight patients, none of whom exhibited a new neurologic deficit, exacerbation of a preexisting neurologic condition, or significant deterioration in neuropsychometric performance. We conclude that the brief arrest of cerebral circulation induced during insertion of an implantable cardioverter-defibrillator is not associated with permanent neurologic injury. (J THORAC CARDIOVASC SURG 1995; 109: 565-73).

AB - During placement of implantable cardioverter-defibrillators, ventricular arrhythmias are induced to test the function of the devices. Although cerebral hypoperfusion and ischemic electroencephalographic changes occur in patients while implantable cardioverter-defibrillators are being tested, no investigation has assessed neurologic outcome in these patients. Nine patients having either implantation or change of an implantable cardioverter-defibrillator underwent neurologic examination and neuropsychometric tests before and after the operation. After induction of general anesthesia and insertion of implantable cardioverter-defibrillator leads (when needed), ventricular fibrillation, ventricular flutter, or ventricular tachycardia, was induced by means of programmed electrical stimulation. Implantable cardioverter-defibrillator testing continued until satisfactory lead placement was confirmed. The intraoperative electroencephalographic recording was analyzed for evidence of ischemic change. In all, an electroencephalogram was recorded during 50 periods of circulatory arrest. Mean duration of the arrest periods was 13.6 seconds. By means of conventional visual inspection of the raw electroencephalogram, high-amplitude rhythmic delta or theta, voltage attenuation, or loss of fast frequency activity was observed in 30 of the arrests. By means of an automated technique of electroencephalographic interpretation based on power spectral analysis, electroencephalographic changes were correctly identified in 26 of the arrests. The incidence of these electroencephalographic changes was dependent on the arrest duration. The mean interval from arrest onset to electroencephalographic change was 7.5 seconds (standard deviation ± 1.8 seconds). In patients with electroencephalographic changes during multiple arrests, no downward trend in this interval was detected in later arrests and no evidence of persistent ischemic change was observed in electroencephalograms recorded after the conclusion of implantable cardioverter-defibrillator testing. Postoperative neurologic and neuropsychometric testing was completed in eight patients, none of whom exhibited a new neurologic deficit, exacerbation of a preexisting neurologic condition, or significant deterioration in neuropsychometric performance. We conclude that the brief arrest of cerebral circulation induced during insertion of an implantable cardioverter-defibrillator is not associated with permanent neurologic injury. (J THORAC CARDIOVASC SURG 1995; 109: 565-73).

UR - http://www.scopus.com/inward/record.url?scp=0028955295&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0028955295&partnerID=8YFLogxK

U2 - 10.1016/S0022-5223(95)70290-3

DO - 10.1016/S0022-5223(95)70290-3

M3 - Article

VL - 109

SP - 565

EP - 573

JO - Journal of Thoracic and Cardiovascular Surgery

JF - Journal of Thoracic and Cardiovascular Surgery

SN - 0022-5223

IS - 3

ER -