EQPlanar

A maximum-likelihood method for accurate organ activity estimation from whole body planar projections

Na Song, B. He, R. L. Wahl, E. C. Frey

Research output: Contribution to journalArticle

6 Citations (Scopus)

Abstract

Optimizing targeted radionuclide therapy requires patient-specific estimation of organ doses. The organ doses are estimated from quantitative nuclear medicine imaging studies, many of which involve planar whole body scans. We have previously developed the quantitative planar (QPlanar) processing method and demonstrated its ability to provide more accurate activity estimates than conventional geometric-mean-based planar (CPlanar) processing methods using physical phantom and simulation studies. The QPlanar method uses the maximum likelihood-expectation maximization algorithm, 3D organ volume of interests (VOIs), and rigorous models of physical image degrading factors to estimate organ activities. However, the QPlanar method requires alignment between the 3D organ VOIs and the 2D planar projections and assumes uniform activity distribution in each VOI. This makes application to patients challenging. As a result, in this paper we propose an extended QPlanar (EQPlanar) method that provides independent-organ rigid registration and includes multiple background regions. We have validated this method using both Monte Carlo simulation and patient data. In the simulation study, we evaluated the precision and accuracy of the method in comparison to the original QPlanar method. For the patient studies, we compared organ activity estimates at 24 h after injection with those from conventional geometric mean-based planar quantification using a 24 h post-injection quantitative SPECT reconstruction as the gold standard. We also compared the goodness of fit of the measured and estimated projections obtained from the EQPlanar method to those from the original method at four other time points where gold standard data were not available. In the simulation study, more accurate activity estimates were provided by the EQPlanar method for all the organs at all the time points compared with the QPlanar method. Based on the patient data, we concluded that the EQPlanar method provided a substantial increase in accuracy of organ activity estimates from 24 h planar images compared to the CPlanar using 24 h SPECT as the golden standard. For other time points, where no golden standard is available, better agreement between estimated and measured projections was observed by using the EQPlanar method compared to the QPlanar method. This phenomenon is consistent with the improvement in goodness of fit seen in both simulation data and 24 h patient data. Therefore, this indicates the improved reliability of organ activity estimates obtained though the EQPlanar method.

Original languageEnglish (US)
Pages (from-to)5503-5524
Number of pages22
JournalPhysics in Medicine and Biology
Volume56
Issue number17
DOIs
StatePublished - Sep 7 2011
Externally publishedYes

Fingerprint

Organ Size
Single-Photon Emission-Computed Tomography
Patient Simulation
Whole Body Imaging
Injections
Nuclear Medicine
Radioisotopes
Therapeutics

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

EQPlanar : A maximum-likelihood method for accurate organ activity estimation from whole body planar projections. / Song, Na; He, B.; Wahl, R. L.; Frey, E. C.

In: Physics in Medicine and Biology, Vol. 56, No. 17, 07.09.2011, p. 5503-5524.

Research output: Contribution to journalArticle

@article{a5996dedd9dc4a519bd8f8eedb8a8ccc,
title = "EQPlanar: A maximum-likelihood method for accurate organ activity estimation from whole body planar projections",
abstract = "Optimizing targeted radionuclide therapy requires patient-specific estimation of organ doses. The organ doses are estimated from quantitative nuclear medicine imaging studies, many of which involve planar whole body scans. We have previously developed the quantitative planar (QPlanar) processing method and demonstrated its ability to provide more accurate activity estimates than conventional geometric-mean-based planar (CPlanar) processing methods using physical phantom and simulation studies. The QPlanar method uses the maximum likelihood-expectation maximization algorithm, 3D organ volume of interests (VOIs), and rigorous models of physical image degrading factors to estimate organ activities. However, the QPlanar method requires alignment between the 3D organ VOIs and the 2D planar projections and assumes uniform activity distribution in each VOI. This makes application to patients challenging. As a result, in this paper we propose an extended QPlanar (EQPlanar) method that provides independent-organ rigid registration and includes multiple background regions. We have validated this method using both Monte Carlo simulation and patient data. In the simulation study, we evaluated the precision and accuracy of the method in comparison to the original QPlanar method. For the patient studies, we compared organ activity estimates at 24 h after injection with those from conventional geometric mean-based planar quantification using a 24 h post-injection quantitative SPECT reconstruction as the gold standard. We also compared the goodness of fit of the measured and estimated projections obtained from the EQPlanar method to those from the original method at four other time points where gold standard data were not available. In the simulation study, more accurate activity estimates were provided by the EQPlanar method for all the organs at all the time points compared with the QPlanar method. Based on the patient data, we concluded that the EQPlanar method provided a substantial increase in accuracy of organ activity estimates from 24 h planar images compared to the CPlanar using 24 h SPECT as the golden standard. For other time points, where no golden standard is available, better agreement between estimated and measured projections was observed by using the EQPlanar method compared to the QPlanar method. This phenomenon is consistent with the improvement in goodness of fit seen in both simulation data and 24 h patient data. Therefore, this indicates the improved reliability of organ activity estimates obtained though the EQPlanar method.",
author = "Na Song and B. He and Wahl, {R. L.} and Frey, {E. C.}",
year = "2011",
month = "9",
day = "7",
doi = "10.1088/0031-9155/56/17/004",
language = "English (US)",
volume = "56",
pages = "5503--5524",
journal = "Physics in Medicine and Biology",
issn = "0031-9155",
publisher = "IOP Publishing Ltd.",
number = "17",

}

TY - JOUR

T1 - EQPlanar

T2 - A maximum-likelihood method for accurate organ activity estimation from whole body planar projections

AU - Song, Na

AU - He, B.

AU - Wahl, R. L.

AU - Frey, E. C.

PY - 2011/9/7

Y1 - 2011/9/7

N2 - Optimizing targeted radionuclide therapy requires patient-specific estimation of organ doses. The organ doses are estimated from quantitative nuclear medicine imaging studies, many of which involve planar whole body scans. We have previously developed the quantitative planar (QPlanar) processing method and demonstrated its ability to provide more accurate activity estimates than conventional geometric-mean-based planar (CPlanar) processing methods using physical phantom and simulation studies. The QPlanar method uses the maximum likelihood-expectation maximization algorithm, 3D organ volume of interests (VOIs), and rigorous models of physical image degrading factors to estimate organ activities. However, the QPlanar method requires alignment between the 3D organ VOIs and the 2D planar projections and assumes uniform activity distribution in each VOI. This makes application to patients challenging. As a result, in this paper we propose an extended QPlanar (EQPlanar) method that provides independent-organ rigid registration and includes multiple background regions. We have validated this method using both Monte Carlo simulation and patient data. In the simulation study, we evaluated the precision and accuracy of the method in comparison to the original QPlanar method. For the patient studies, we compared organ activity estimates at 24 h after injection with those from conventional geometric mean-based planar quantification using a 24 h post-injection quantitative SPECT reconstruction as the gold standard. We also compared the goodness of fit of the measured and estimated projections obtained from the EQPlanar method to those from the original method at four other time points where gold standard data were not available. In the simulation study, more accurate activity estimates were provided by the EQPlanar method for all the organs at all the time points compared with the QPlanar method. Based on the patient data, we concluded that the EQPlanar method provided a substantial increase in accuracy of organ activity estimates from 24 h planar images compared to the CPlanar using 24 h SPECT as the golden standard. For other time points, where no golden standard is available, better agreement between estimated and measured projections was observed by using the EQPlanar method compared to the QPlanar method. This phenomenon is consistent with the improvement in goodness of fit seen in both simulation data and 24 h patient data. Therefore, this indicates the improved reliability of organ activity estimates obtained though the EQPlanar method.

AB - Optimizing targeted radionuclide therapy requires patient-specific estimation of organ doses. The organ doses are estimated from quantitative nuclear medicine imaging studies, many of which involve planar whole body scans. We have previously developed the quantitative planar (QPlanar) processing method and demonstrated its ability to provide more accurate activity estimates than conventional geometric-mean-based planar (CPlanar) processing methods using physical phantom and simulation studies. The QPlanar method uses the maximum likelihood-expectation maximization algorithm, 3D organ volume of interests (VOIs), and rigorous models of physical image degrading factors to estimate organ activities. However, the QPlanar method requires alignment between the 3D organ VOIs and the 2D planar projections and assumes uniform activity distribution in each VOI. This makes application to patients challenging. As a result, in this paper we propose an extended QPlanar (EQPlanar) method that provides independent-organ rigid registration and includes multiple background regions. We have validated this method using both Monte Carlo simulation and patient data. In the simulation study, we evaluated the precision and accuracy of the method in comparison to the original QPlanar method. For the patient studies, we compared organ activity estimates at 24 h after injection with those from conventional geometric mean-based planar quantification using a 24 h post-injection quantitative SPECT reconstruction as the gold standard. We also compared the goodness of fit of the measured and estimated projections obtained from the EQPlanar method to those from the original method at four other time points where gold standard data were not available. In the simulation study, more accurate activity estimates were provided by the EQPlanar method for all the organs at all the time points compared with the QPlanar method. Based on the patient data, we concluded that the EQPlanar method provided a substantial increase in accuracy of organ activity estimates from 24 h planar images compared to the CPlanar using 24 h SPECT as the golden standard. For other time points, where no golden standard is available, better agreement between estimated and measured projections was observed by using the EQPlanar method compared to the QPlanar method. This phenomenon is consistent with the improvement in goodness of fit seen in both simulation data and 24 h patient data. Therefore, this indicates the improved reliability of organ activity estimates obtained though the EQPlanar method.

UR - http://www.scopus.com/inward/record.url?scp=80052053917&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=80052053917&partnerID=8YFLogxK

U2 - 10.1088/0031-9155/56/17/004

DO - 10.1088/0031-9155/56/17/004

M3 - Article

VL - 56

SP - 5503

EP - 5524

JO - Physics in Medicine and Biology

JF - Physics in Medicine and Biology

SN - 0031-9155

IS - 17

ER -