TY - JOUR
T1 - Endless Loop Tachycardia in an AV Universal (DDD) Pacemaker
AU - FURMAN, SEYMOUR
AU - FISHER, JOHN D.
PY - 1982/7
Y1 - 1982/7
N2 - Implantation of an A V universal (DDD) pacemaker, stimulating and sensing both atrium and ventricle caused a pacemaker mediated tachycardia, begun by a P wave occurring after a premature ventricular contraction, and sustained by P waves caused by retrograde AV conduction. The tachycardia was terminated by application of a magnet to the pulse generator which temporarily made it insensitive to incoming signals, both P and QRS. (PACE, Vol. 5. July‐August, 1982)
AB - Implantation of an A V universal (DDD) pacemaker, stimulating and sensing both atrium and ventricle caused a pacemaker mediated tachycardia, begun by a P wave occurring after a premature ventricular contraction, and sustained by P waves caused by retrograde AV conduction. The tachycardia was terminated by application of a magnet to the pulse generator which temporarily made it insensitive to incoming signals, both P and QRS. (PACE, Vol. 5. July‐August, 1982)
UR - http://www.scopus.com/inward/record.url?scp=0019966637&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0019966637&partnerID=8YFLogxK
U2 - 10.1111/j.1540-8159.1982.tb02265.x
DO - 10.1111/j.1540-8159.1982.tb02265.x
M3 - Article
C2 - 6180386
AN - SCOPUS:0019966637
SN - 0147-8389
VL - 5
SP - 486
EP - 489
JO - Pacing and Clinical Electrophysiology
JF - Pacing and Clinical Electrophysiology
IS - 4
ER -