Dosimetric characteristics of a double-focused miniature multileaf collimator

Sanford L. Meeks, Francis J. Bova, Siyong Kim, Wolfgang A. Tome, John M. Buatti, William A. Friedman

Research output: Contribution to journalArticle

33 Citations (Scopus)

Abstract

The dosimetric characteristics of a double-focused miniature multileaf collimator (mMLC) attached to a Philips SL75/5 linear accelerator (linac) have been investigated. Output factors, percentage depth-dose, penumbra, leaf transmission, and leakage between the leaves were measured for the 6 MV x- ray beam on this accelerator. Because leakage both through and between the leaves is minimal, the linac jaws can be kept fixed while the mMLC leaf configuration is modified for different aperture shapes. This allows for accurate output prediction using the equivalent square formalism. Percent depth-dose measured for fields defined by the mMLC show little deviation from the percent depth-dose measured for fields defined by the machine jaws or Lipowitz metal blocks. Because the mMLC matches beam divergence in both directions, allows minimal beam transmission, and has a large source-to- collimator distance, the penumbra is sharper for fields defined by the mMLC than for fields defined by the linac jaws or Lipowitz metal blocks. Based on these data, dose calculations for mMLC-defined fields can be applied with no change in procedures from those used for fields defined using conventional methods.

Original languageEnglish (US)
Pages (from-to)729-733
Number of pages5
JournalMedical Physics
Volume26
Issue number5
DOIs
StatePublished - May 1999
Externally publishedYes

Fingerprint

Particle Accelerators
Jaw
Metals
X-Rays

Keywords

  • Multileaf collimator
  • X-ray dosimetry

ASJC Scopus subject areas

  • Biophysics

Cite this

Meeks, S. L., Bova, F. J., Kim, S., Tome, W. A., Buatti, J. M., & Friedman, W. A. (1999). Dosimetric characteristics of a double-focused miniature multileaf collimator. Medical Physics, 26(5), 729-733. https://doi.org/10.1118/1.598579

Dosimetric characteristics of a double-focused miniature multileaf collimator. / Meeks, Sanford L.; Bova, Francis J.; Kim, Siyong; Tome, Wolfgang A.; Buatti, John M.; Friedman, William A.

In: Medical Physics, Vol. 26, No. 5, 05.1999, p. 729-733.

Research output: Contribution to journalArticle

Meeks, SL, Bova, FJ, Kim, S, Tome, WA, Buatti, JM & Friedman, WA 1999, 'Dosimetric characteristics of a double-focused miniature multileaf collimator', Medical Physics, vol. 26, no. 5, pp. 729-733. https://doi.org/10.1118/1.598579
Meeks, Sanford L. ; Bova, Francis J. ; Kim, Siyong ; Tome, Wolfgang A. ; Buatti, John M. ; Friedman, William A. / Dosimetric characteristics of a double-focused miniature multileaf collimator. In: Medical Physics. 1999 ; Vol. 26, No. 5. pp. 729-733.
@article{7b39b5d5cd21486eae1ea39a76c53469,
title = "Dosimetric characteristics of a double-focused miniature multileaf collimator",
abstract = "The dosimetric characteristics of a double-focused miniature multileaf collimator (mMLC) attached to a Philips SL75/5 linear accelerator (linac) have been investigated. Output factors, percentage depth-dose, penumbra, leaf transmission, and leakage between the leaves were measured for the 6 MV x- ray beam on this accelerator. Because leakage both through and between the leaves is minimal, the linac jaws can be kept fixed while the mMLC leaf configuration is modified for different aperture shapes. This allows for accurate output prediction using the equivalent square formalism. Percent depth-dose measured for fields defined by the mMLC show little deviation from the percent depth-dose measured for fields defined by the machine jaws or Lipowitz metal blocks. Because the mMLC matches beam divergence in both directions, allows minimal beam transmission, and has a large source-to- collimator distance, the penumbra is sharper for fields defined by the mMLC than for fields defined by the linac jaws or Lipowitz metal blocks. Based on these data, dose calculations for mMLC-defined fields can be applied with no change in procedures from those used for fields defined using conventional methods.",
keywords = "Multileaf collimator, X-ray dosimetry",
author = "Meeks, {Sanford L.} and Bova, {Francis J.} and Siyong Kim and Tome, {Wolfgang A.} and Buatti, {John M.} and Friedman, {William A.}",
year = "1999",
month = "5",
doi = "10.1118/1.598579",
language = "English (US)",
volume = "26",
pages = "729--733",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "5",

}

TY - JOUR

T1 - Dosimetric characteristics of a double-focused miniature multileaf collimator

AU - Meeks, Sanford L.

AU - Bova, Francis J.

AU - Kim, Siyong

AU - Tome, Wolfgang A.

AU - Buatti, John M.

AU - Friedman, William A.

PY - 1999/5

Y1 - 1999/5

N2 - The dosimetric characteristics of a double-focused miniature multileaf collimator (mMLC) attached to a Philips SL75/5 linear accelerator (linac) have been investigated. Output factors, percentage depth-dose, penumbra, leaf transmission, and leakage between the leaves were measured for the 6 MV x- ray beam on this accelerator. Because leakage both through and between the leaves is minimal, the linac jaws can be kept fixed while the mMLC leaf configuration is modified for different aperture shapes. This allows for accurate output prediction using the equivalent square formalism. Percent depth-dose measured for fields defined by the mMLC show little deviation from the percent depth-dose measured for fields defined by the machine jaws or Lipowitz metal blocks. Because the mMLC matches beam divergence in both directions, allows minimal beam transmission, and has a large source-to- collimator distance, the penumbra is sharper for fields defined by the mMLC than for fields defined by the linac jaws or Lipowitz metal blocks. Based on these data, dose calculations for mMLC-defined fields can be applied with no change in procedures from those used for fields defined using conventional methods.

AB - The dosimetric characteristics of a double-focused miniature multileaf collimator (mMLC) attached to a Philips SL75/5 linear accelerator (linac) have been investigated. Output factors, percentage depth-dose, penumbra, leaf transmission, and leakage between the leaves were measured for the 6 MV x- ray beam on this accelerator. Because leakage both through and between the leaves is minimal, the linac jaws can be kept fixed while the mMLC leaf configuration is modified for different aperture shapes. This allows for accurate output prediction using the equivalent square formalism. Percent depth-dose measured for fields defined by the mMLC show little deviation from the percent depth-dose measured for fields defined by the machine jaws or Lipowitz metal blocks. Because the mMLC matches beam divergence in both directions, allows minimal beam transmission, and has a large source-to- collimator distance, the penumbra is sharper for fields defined by the mMLC than for fields defined by the linac jaws or Lipowitz metal blocks. Based on these data, dose calculations for mMLC-defined fields can be applied with no change in procedures from those used for fields defined using conventional methods.

KW - Multileaf collimator

KW - X-ray dosimetry

UR - http://www.scopus.com/inward/record.url?scp=0032955179&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0032955179&partnerID=8YFLogxK

U2 - 10.1118/1.598579

DO - 10.1118/1.598579

M3 - Article

C2 - 10360533

AN - SCOPUS:0032955179

VL - 26

SP - 729

EP - 733

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 5

ER -