TY - JOUR
T1 - Clinical Experiences With Onboard Imager KV Images for Linear Accelerator-Based Stereotactic Radiosurgery and Radiotherapy Setup
AU - Hong, Linda X.
AU - Chen, Chin C.
AU - Garg, Madhur
AU - Yaparpalvi, Ravindra
AU - Mah, Dennis
N1 - Copyright:
Copyright 2009 Elsevier B.V., All rights reserved.
PY - 2009/2/1
Y1 - 2009/2/1
N2 - Purpose: To report our clinical experiences with on-board imager (OBI) kV image verification for cranial stereotactic radiosurgery (SRS) and radiotherapy (SRT) treatments. Methods and Materials: Between January 2007 and May 2008, 42 patients (57 lesions) were treated with SRS with head frame immobilization and 13 patients (14 lesions) were treated with SRT with face mask immobilization at our institution. No margin was added to the gross tumor for SRS patients, and a 3-mm three-dimensional margin was added to the gross tumor to create the planning target volume for SRT patients. After localizing the patient with stereotactic target positioner (TaPo), orthogonal kV images using OBI were taken and fused to planning digital reconstructed radiographs. Suggested couch shifts in vertical, longitudinal, and lateral directions were recorded. kV images were also taken immediately after treatment for 21 SRS patients and on a weekly basis for 6 SRT patients to assess any intrafraction changes. Results: For SRS patients, 57 pretreatment kV images were evaluated and the suggested shifts were all within 1 mm in any direction (i.e., within the accuracy of image fusion). For SRT patients, the suggested shifts were out of the 3-mm tolerance for 31 of 309 setups. Intrafraction motions were detected in 3 SRT patients. Conclusions: kV imaging provided a useful tool for SRS or SRT setups. For SRS setup with head frame, it provides radiographic confirmation of localization using the stereotactic target positioner. For SRT with mask, a 3-mm margin is adequate and feasible for routine setup when TaPo is combined with kV imaging.
AB - Purpose: To report our clinical experiences with on-board imager (OBI) kV image verification for cranial stereotactic radiosurgery (SRS) and radiotherapy (SRT) treatments. Methods and Materials: Between January 2007 and May 2008, 42 patients (57 lesions) were treated with SRS with head frame immobilization and 13 patients (14 lesions) were treated with SRT with face mask immobilization at our institution. No margin was added to the gross tumor for SRS patients, and a 3-mm three-dimensional margin was added to the gross tumor to create the planning target volume for SRT patients. After localizing the patient with stereotactic target positioner (TaPo), orthogonal kV images using OBI were taken and fused to planning digital reconstructed radiographs. Suggested couch shifts in vertical, longitudinal, and lateral directions were recorded. kV images were also taken immediately after treatment for 21 SRS patients and on a weekly basis for 6 SRT patients to assess any intrafraction changes. Results: For SRS patients, 57 pretreatment kV images were evaluated and the suggested shifts were all within 1 mm in any direction (i.e., within the accuracy of image fusion). For SRT patients, the suggested shifts were out of the 3-mm tolerance for 31 of 309 setups. Intrafraction motions were detected in 3 SRT patients. Conclusions: kV imaging provided a useful tool for SRS or SRT setups. For SRS setup with head frame, it provides radiographic confirmation of localization using the stereotactic target positioner. For SRT with mask, a 3-mm margin is adequate and feasible for routine setup when TaPo is combined with kV imaging.
KW - LINAC-based stereotactic radiosurgery and radiotherapy
KW - OBI imaging
KW - Setup accuracy
UR - http://www.scopus.com/inward/record.url?scp=58149330212&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=58149330212&partnerID=8YFLogxK
U2 - 10.1016/j.ijrobp.2008.09.055
DO - 10.1016/j.ijrobp.2008.09.055
M3 - Article
C2 - 19147020
AN - SCOPUS:58149330212
SN - 0360-3016
VL - 73
SP - 556
EP - 561
JO - International Journal of Radiation Oncology Biology Physics
JF - International Journal of Radiation Oncology Biology Physics
IS - 2
ER -