Automatic methods for detection of tachyarrhythmias by antitachycardia devices

Frank Pannizzo, Anthony D. Mercando, John Devens Fisher, Seymour Furman

Research output: Contribution to journalArticle

26 Citations (Scopus)

Abstract

Electrical devices play an increasingly important role in the control of tachyarrhythmias. Antitachycardia pacing and automatic defibrillation have been severely limited by the poor specificity of tachycardia discrimination in commercially available devices. Although absolute heart rate has been the principal means of automatic diagnosis, several new detection algorithms and methods are being investigated. Multiple electrode timing comparison, signal processing and pattern recognition are employed in these newer techniques. Although each offers some improvement over present technology, none is capable of identifying all arrhythmias. The methods employing comparison of atrial and ventricular rates, without additional criteria, are unable to detect ventricular tachycardia in the presence of 1:1 retrograde conduction. Electrographic analysis techniques require very stable electrodes and may not tolerate normal morphologic variations. A combination of two or more approaches may ultimately be required. All techniques will require that certain critical variables be programmable to allow for individualization in each clinical situation. Software-controllable devices and those capable of sensing from both the atria and the ventricles will provide the sophistication necessary for the implementation of complex tachycardia detection algorithms. This report reviews automatic tachycardia detection techniques in current use and under investigation.

Original languageEnglish (US)
Pages (from-to)308-316
Number of pages9
JournalJournal of the American College of Cardiology
Volume11
Issue number2
DOIs
StatePublished - 1988

Fingerprint

Tachycardia
Equipment and Supplies
Electrodes
Ventricular Tachycardia
Cardiac Arrhythmias
Software
Heart Rate
Technology

ASJC Scopus subject areas

  • Nursing(all)

Cite this

Automatic methods for detection of tachyarrhythmias by antitachycardia devices. / Pannizzo, Frank; Mercando, Anthony D.; Fisher, John Devens; Furman, Seymour.

In: Journal of the American College of Cardiology, Vol. 11, No. 2, 1988, p. 308-316.

Research output: Contribution to journalArticle

Pannizzo, Frank ; Mercando, Anthony D. ; Fisher, John Devens ; Furman, Seymour. / Automatic methods for detection of tachyarrhythmias by antitachycardia devices. In: Journal of the American College of Cardiology. 1988 ; Vol. 11, No. 2. pp. 308-316.
@article{debdde36dd0549c78cf35a21777eea2b,
title = "Automatic methods for detection of tachyarrhythmias by antitachycardia devices",
abstract = "Electrical devices play an increasingly important role in the control of tachyarrhythmias. Antitachycardia pacing and automatic defibrillation have been severely limited by the poor specificity of tachycardia discrimination in commercially available devices. Although absolute heart rate has been the principal means of automatic diagnosis, several new detection algorithms and methods are being investigated. Multiple electrode timing comparison, signal processing and pattern recognition are employed in these newer techniques. Although each offers some improvement over present technology, none is capable of identifying all arrhythmias. The methods employing comparison of atrial and ventricular rates, without additional criteria, are unable to detect ventricular tachycardia in the presence of 1:1 retrograde conduction. Electrographic analysis techniques require very stable electrodes and may not tolerate normal morphologic variations. A combination of two or more approaches may ultimately be required. All techniques will require that certain critical variables be programmable to allow for individualization in each clinical situation. Software-controllable devices and those capable of sensing from both the atria and the ventricles will provide the sophistication necessary for the implementation of complex tachycardia detection algorithms. This report reviews automatic tachycardia detection techniques in current use and under investigation.",
author = "Frank Pannizzo and Mercando, {Anthony D.} and Fisher, {John Devens} and Seymour Furman",
year = "1988",
doi = "10.1016/0735-1097(88)90095-2",
language = "English (US)",
volume = "11",
pages = "308--316",
journal = "Journal of the American College of Cardiology",
issn = "0735-1097",
publisher = "Elsevier USA",
number = "2",

}

TY - JOUR

T1 - Automatic methods for detection of tachyarrhythmias by antitachycardia devices

AU - Pannizzo, Frank

AU - Mercando, Anthony D.

AU - Fisher, John Devens

AU - Furman, Seymour

PY - 1988

Y1 - 1988

N2 - Electrical devices play an increasingly important role in the control of tachyarrhythmias. Antitachycardia pacing and automatic defibrillation have been severely limited by the poor specificity of tachycardia discrimination in commercially available devices. Although absolute heart rate has been the principal means of automatic diagnosis, several new detection algorithms and methods are being investigated. Multiple electrode timing comparison, signal processing and pattern recognition are employed in these newer techniques. Although each offers some improvement over present technology, none is capable of identifying all arrhythmias. The methods employing comparison of atrial and ventricular rates, without additional criteria, are unable to detect ventricular tachycardia in the presence of 1:1 retrograde conduction. Electrographic analysis techniques require very stable electrodes and may not tolerate normal morphologic variations. A combination of two or more approaches may ultimately be required. All techniques will require that certain critical variables be programmable to allow for individualization in each clinical situation. Software-controllable devices and those capable of sensing from both the atria and the ventricles will provide the sophistication necessary for the implementation of complex tachycardia detection algorithms. This report reviews automatic tachycardia detection techniques in current use and under investigation.

AB - Electrical devices play an increasingly important role in the control of tachyarrhythmias. Antitachycardia pacing and automatic defibrillation have been severely limited by the poor specificity of tachycardia discrimination in commercially available devices. Although absolute heart rate has been the principal means of automatic diagnosis, several new detection algorithms and methods are being investigated. Multiple electrode timing comparison, signal processing and pattern recognition are employed in these newer techniques. Although each offers some improvement over present technology, none is capable of identifying all arrhythmias. The methods employing comparison of atrial and ventricular rates, without additional criteria, are unable to detect ventricular tachycardia in the presence of 1:1 retrograde conduction. Electrographic analysis techniques require very stable electrodes and may not tolerate normal morphologic variations. A combination of two or more approaches may ultimately be required. All techniques will require that certain critical variables be programmable to allow for individualization in each clinical situation. Software-controllable devices and those capable of sensing from both the atria and the ventricles will provide the sophistication necessary for the implementation of complex tachycardia detection algorithms. This report reviews automatic tachycardia detection techniques in current use and under investigation.

UR - http://www.scopus.com/inward/record.url?scp=0023911169&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0023911169&partnerID=8YFLogxK

U2 - 10.1016/0735-1097(88)90095-2

DO - 10.1016/0735-1097(88)90095-2

M3 - Article

C2 - 3276754

AN - SCOPUS:0023911169

VL - 11

SP - 308

EP - 316

JO - Journal of the American College of Cardiology

JF - Journal of the American College of Cardiology

SN - 0735-1097

IS - 2