A gamma camera count rate saturation correction method for whole-body planar imaging

Robert F. Hobbs, Sébastien Baechler, Srinivasan Senthamizhchelvan, Andrew R. Prideaux, Caroline E. Esaias, Melvin Reinhardt, Eric C. Frey, David M. Loeb, George Sgouros

Research output: Contribution to journalArticle

8 Citations (Scopus)

Abstract

Whole-body (WB) planar imaging has long been one of the staple methods of dosimetry, and its quantification has been formalized by the MIRD Committee in pamphlet no 16. One of the issues not specifically addressed in the formalism occurs when the count rates reaching the detector are sufficiently high to result in camera count saturation. Camera dead-time effects have been extensively studied, but all of the developed correction methods assume static acquisitions. However, during WB planar (sweep) imaging, a variable amount of imaged activity exists in the detector's field of view as a function of time and therefore the camera saturation is time dependent. A new time-dependent algorithm was developed to correct for dead-time effects during WB planar acquisitions that accounts for relative motion between detector heads and imaged object. Static camera dead-time parameters were acquired by imaging decaying activity in a phantom and obtaining a saturation curve. Using these parameters, an iterative algorithm akin to Newton's method was developed, which takes into account the variable count rate seen by the detector as a function of time. The algorithm was tested on simulated data as well as on a whole-body scan of high activity Samarium-153 in an ellipsoid phantom. A complete set of parameters from unsaturated phantom data necessary for count rate to activity conversion was also obtained, including build-up and attenuation coefficients, in order to convert corrected count rate values to activity. The algorithm proved successful in accounting for motion- and time-dependent saturation effects in both the simulated and measured data and converged to any desired degree of precision. The clearance half-life calculated from the ellipsoid phantom data was calculated to be 45.1 h after dead-time correction and 51.4 h with no correction; the physical decay half-life of Samarium-153 is 46.3 h. Accurate WB planar dosimetry of high activities relies on successfully compensating for camera saturation which takes into account the variable activity in the field of view, i.e. time-dependent dead-time effects. The algorithm presented here accomplishes this task.

Original languageEnglish (US)
Pages (from-to)817-821
Number of pages5
JournalPhysics in Medicine and Biology
Volume55
Issue number3
DOIs
StatePublished - Feb 12 2010
Externally publishedYes

Fingerprint

Whole Body Imaging
Gamma Cameras
Samarium
Half-Life
Pamphlets
Head

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

Hobbs, R. F., Baechler, S., Senthamizhchelvan, S., Prideaux, A. R., Esaias, C. E., Reinhardt, M., ... Sgouros, G. (2010). A gamma camera count rate saturation correction method for whole-body planar imaging. Physics in Medicine and Biology, 55(3), 817-821. https://doi.org/10.1088/0031-9155/55/3/018

A gamma camera count rate saturation correction method for whole-body planar imaging. / Hobbs, Robert F.; Baechler, Sébastien; Senthamizhchelvan, Srinivasan; Prideaux, Andrew R.; Esaias, Caroline E.; Reinhardt, Melvin; Frey, Eric C.; Loeb, David M.; Sgouros, George.

In: Physics in Medicine and Biology, Vol. 55, No. 3, 12.02.2010, p. 817-821.

Research output: Contribution to journalArticle

Hobbs, RF, Baechler, S, Senthamizhchelvan, S, Prideaux, AR, Esaias, CE, Reinhardt, M, Frey, EC, Loeb, DM & Sgouros, G 2010, 'A gamma camera count rate saturation correction method for whole-body planar imaging', Physics in Medicine and Biology, vol. 55, no. 3, pp. 817-821. https://doi.org/10.1088/0031-9155/55/3/018
Hobbs RF, Baechler S, Senthamizhchelvan S, Prideaux AR, Esaias CE, Reinhardt M et al. A gamma camera count rate saturation correction method for whole-body planar imaging. Physics in Medicine and Biology. 2010 Feb 12;55(3):817-821. https://doi.org/10.1088/0031-9155/55/3/018
Hobbs, Robert F. ; Baechler, Sébastien ; Senthamizhchelvan, Srinivasan ; Prideaux, Andrew R. ; Esaias, Caroline E. ; Reinhardt, Melvin ; Frey, Eric C. ; Loeb, David M. ; Sgouros, George. / A gamma camera count rate saturation correction method for whole-body planar imaging. In: Physics in Medicine and Biology. 2010 ; Vol. 55, No. 3. pp. 817-821.
@article{b0da4179db7f47cf89fdac296363871e,
title = "A gamma camera count rate saturation correction method for whole-body planar imaging",
abstract = "Whole-body (WB) planar imaging has long been one of the staple methods of dosimetry, and its quantification has been formalized by the MIRD Committee in pamphlet no 16. One of the issues not specifically addressed in the formalism occurs when the count rates reaching the detector are sufficiently high to result in camera count saturation. Camera dead-time effects have been extensively studied, but all of the developed correction methods assume static acquisitions. However, during WB planar (sweep) imaging, a variable amount of imaged activity exists in the detector's field of view as a function of time and therefore the camera saturation is time dependent. A new time-dependent algorithm was developed to correct for dead-time effects during WB planar acquisitions that accounts for relative motion between detector heads and imaged object. Static camera dead-time parameters were acquired by imaging decaying activity in a phantom and obtaining a saturation curve. Using these parameters, an iterative algorithm akin to Newton's method was developed, which takes into account the variable count rate seen by the detector as a function of time. The algorithm was tested on simulated data as well as on a whole-body scan of high activity Samarium-153 in an ellipsoid phantom. A complete set of parameters from unsaturated phantom data necessary for count rate to activity conversion was also obtained, including build-up and attenuation coefficients, in order to convert corrected count rate values to activity. The algorithm proved successful in accounting for motion- and time-dependent saturation effects in both the simulated and measured data and converged to any desired degree of precision. The clearance half-life calculated from the ellipsoid phantom data was calculated to be 45.1 h after dead-time correction and 51.4 h with no correction; the physical decay half-life of Samarium-153 is 46.3 h. Accurate WB planar dosimetry of high activities relies on successfully compensating for camera saturation which takes into account the variable activity in the field of view, i.e. time-dependent dead-time effects. The algorithm presented here accomplishes this task.",
author = "Hobbs, {Robert F.} and S{\'e}bastien Baechler and Srinivasan Senthamizhchelvan and Prideaux, {Andrew R.} and Esaias, {Caroline E.} and Melvin Reinhardt and Frey, {Eric C.} and Loeb, {David M.} and George Sgouros",
year = "2010",
month = "2",
day = "12",
doi = "10.1088/0031-9155/55/3/018",
language = "English (US)",
volume = "55",
pages = "817--821",
journal = "Physics in Medicine and Biology",
issn = "0031-9155",
publisher = "IOP Publishing Ltd.",
number = "3",

}

TY - JOUR

T1 - A gamma camera count rate saturation correction method for whole-body planar imaging

AU - Hobbs, Robert F.

AU - Baechler, Sébastien

AU - Senthamizhchelvan, Srinivasan

AU - Prideaux, Andrew R.

AU - Esaias, Caroline E.

AU - Reinhardt, Melvin

AU - Frey, Eric C.

AU - Loeb, David M.

AU - Sgouros, George

PY - 2010/2/12

Y1 - 2010/2/12

N2 - Whole-body (WB) planar imaging has long been one of the staple methods of dosimetry, and its quantification has been formalized by the MIRD Committee in pamphlet no 16. One of the issues not specifically addressed in the formalism occurs when the count rates reaching the detector are sufficiently high to result in camera count saturation. Camera dead-time effects have been extensively studied, but all of the developed correction methods assume static acquisitions. However, during WB planar (sweep) imaging, a variable amount of imaged activity exists in the detector's field of view as a function of time and therefore the camera saturation is time dependent. A new time-dependent algorithm was developed to correct for dead-time effects during WB planar acquisitions that accounts for relative motion between detector heads and imaged object. Static camera dead-time parameters were acquired by imaging decaying activity in a phantom and obtaining a saturation curve. Using these parameters, an iterative algorithm akin to Newton's method was developed, which takes into account the variable count rate seen by the detector as a function of time. The algorithm was tested on simulated data as well as on a whole-body scan of high activity Samarium-153 in an ellipsoid phantom. A complete set of parameters from unsaturated phantom data necessary for count rate to activity conversion was also obtained, including build-up and attenuation coefficients, in order to convert corrected count rate values to activity. The algorithm proved successful in accounting for motion- and time-dependent saturation effects in both the simulated and measured data and converged to any desired degree of precision. The clearance half-life calculated from the ellipsoid phantom data was calculated to be 45.1 h after dead-time correction and 51.4 h with no correction; the physical decay half-life of Samarium-153 is 46.3 h. Accurate WB planar dosimetry of high activities relies on successfully compensating for camera saturation which takes into account the variable activity in the field of view, i.e. time-dependent dead-time effects. The algorithm presented here accomplishes this task.

AB - Whole-body (WB) planar imaging has long been one of the staple methods of dosimetry, and its quantification has been formalized by the MIRD Committee in pamphlet no 16. One of the issues not specifically addressed in the formalism occurs when the count rates reaching the detector are sufficiently high to result in camera count saturation. Camera dead-time effects have been extensively studied, but all of the developed correction methods assume static acquisitions. However, during WB planar (sweep) imaging, a variable amount of imaged activity exists in the detector's field of view as a function of time and therefore the camera saturation is time dependent. A new time-dependent algorithm was developed to correct for dead-time effects during WB planar acquisitions that accounts for relative motion between detector heads and imaged object. Static camera dead-time parameters were acquired by imaging decaying activity in a phantom and obtaining a saturation curve. Using these parameters, an iterative algorithm akin to Newton's method was developed, which takes into account the variable count rate seen by the detector as a function of time. The algorithm was tested on simulated data as well as on a whole-body scan of high activity Samarium-153 in an ellipsoid phantom. A complete set of parameters from unsaturated phantom data necessary for count rate to activity conversion was also obtained, including build-up and attenuation coefficients, in order to convert corrected count rate values to activity. The algorithm proved successful in accounting for motion- and time-dependent saturation effects in both the simulated and measured data and converged to any desired degree of precision. The clearance half-life calculated from the ellipsoid phantom data was calculated to be 45.1 h after dead-time correction and 51.4 h with no correction; the physical decay half-life of Samarium-153 is 46.3 h. Accurate WB planar dosimetry of high activities relies on successfully compensating for camera saturation which takes into account the variable activity in the field of view, i.e. time-dependent dead-time effects. The algorithm presented here accomplishes this task.

UR - http://www.scopus.com/inward/record.url?scp=76149137458&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=76149137458&partnerID=8YFLogxK

U2 - 10.1088/0031-9155/55/3/018

DO - 10.1088/0031-9155/55/3/018

M3 - Article

C2 - 20071766

AN - SCOPUS:76149137458

VL - 55

SP - 817

EP - 821

JO - Physics in Medicine and Biology

JF - Physics in Medicine and Biology

SN - 0031-9155

IS - 3

ER -